Laboratory Medicine Department, La Paz University Hospital, Madrid, Spain.
Clinical Biochemistry Department, Hospital Beaujon, Clichy, France.
Biochem Med (Zagreb). 2022 Jun 15;32(2):020709. doi: 10.11613/BM.2022.020709.
The Fourth Universal Definition of Myocardial Infarction Global Taskforce recommends the use of high sensitive troponin (hs-Tn) assays in the diagnosis of acute myocardial infarction. We evaluated the analytical performance of the Atellica IM High-sensitivity Troponin I Assay (hs-TnI) (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) and compared its performance to other hs-TnI assays (Siemens Advia Centaur, Dimension Vista, Dimension EXL, and Abbott Architect (Wiesbaden, Germany)) at one or more sites across Europe.
Precision, detection limit, linearity, method comparison, and interference studies were performed according to Clinical and Laboratory Standards Institute protocols. Values in 40 healthy individuals were compared to the manufacturer's cut-offs. Sample turnaround time (TAT) was examined.
Imprecision repeatability CVs were 1.1-4.7% and within-lab imprecision were 1.8-7.6% (10.0-25,000 ng/L). The limit of blank (LoB), detection (LoD), and quantitation (LoQ) aligned with the manufacturer's values of 0.5 ng/L, 1.6 ng/L, and 2.5 ng/L, respectively. Passing-Bablok regression demonstrated good correlations between Atellica IM analyser with other systems; some minor deviations were observed. All results in healthy volunteers fell below the 99th percentile URL, and greater than 50% of each sex demonstrated values above the LoD. No interference was observed for biotin (≤ 1500 µg/L), but a slight bias at 5.0 g/L haemoglobin and 50 ng/L Tn was observed. TAT from was fast (mean time = 10.9 minutes) and reproducible (6%CV).
Real-world analytical and TAT performance of the hs-TnI assay on the Atellica IM analyser make this assay fit for routine use in clinical laboratories.
第四版心肌梗死全球定义工作组建议在急性心肌梗死的诊断中使用高敏肌钙蛋白(hs-Tn)检测。我们评估了 Atellica IM 高敏肌钙蛋白 I 检测(hs-TnI)(西门子医疗诊断公司,美国塔瑞镇)的分析性能,并在欧洲的一个或多个地点比较了其与其他 hs-TnI 检测(西门子 Advia Centaur、Dimension Vista、Dimension EXL 和 Abbott Architect(德国威斯巴登))的性能。
根据临床和实验室标准协会的方案进行精密度、检测限、线性、方法比较和干扰研究。将 40 名健康个体的值与制造商的截止值进行比较。检查样本周转时间(TAT)。
重复性精密度 CV 为 1.1-4.7%,实验室内精密度为 1.8-7.6%(10.0-25,000ng/L)。空白限(LoB)、检测限(LoD)和定量限(LoQ)分别与制造商规定的 0.5ng/L、1.6ng/L 和 2.5ng/L 一致。Passing-Bablok 回归显示 Atellica IM 分析仪与其他系统之间具有良好的相关性;观察到一些轻微的偏差。健康志愿者的所有结果均低于 99% URL,且超过 50%的男性和女性的结果均高于 LoD。生物素(≤1500μg/L)无干扰,但血红蛋白 5.0g/L 和 Tn 50ng/L 时观察到轻微偏差。TAT 时间较短(平均时间=10.9 分钟)且重现性好(6%CV)。
Atellica IM 分析仪上 hs-TnI 检测的实际分析和 TAT 性能使其适合于临床实验室的常规使用。