自膨胀式经导管主动脉瓣膜的梯度及压力恢复取决于升主动脉大小:一项体外研究

Gradient and pressure recovery of a self-expandable transcatheter aortic valve depends on ascending aorta size: In vitro study.

作者信息

Samaee Milad, Hatoum Hoda, Biersmith Michael, Yeats Breandan, Gooden Shelley C, Thourani Vinod H, Hahn Rebecca T, Lilly Scott, Yoganathan Ajit, Dasi Lakshmi Prasad

机构信息

Department of Biomedical Engineering, Georgia Institute of Technology, Atlanta, Ga.

Department of Biomedical Engineering, Michigan Technological University, Houghton, Mich.

出版信息

JTCVS Open. 2022 Jan 22;9:28-38. doi: 10.1016/j.xjon.2022.01.003. eCollection 2022 Mar.

Abstract

OBJECTIVE

In this study we aimed to understand the role of interaction of the Medtronic Evolut R transcatheter aortic valve with the ascending aorta (AA) by evaluating the performance of the valve and the pressure recovery in different AA diameters with the same aortic annulus size.

METHODS

A 26-mm Medtronic Evolut R valve was tested using a left heart simulator in aortic root models of different AA diameter (D): small (D = 23 mm), medium (D = 28 mm), and large (D = 34 mm) under physiological conditions. Measurements of pressure from upstream to downstream of the valve were performed using a catheter at small intervals to comprehensively assess pressure gradient and pressure recovery.

RESULTS

In the small AA, the measured peak and mean pressure gradient at vena contracta were 11.5 ± 0.5 mm Hg and 7.8 ± 0.4 mm Hg, respectively, which was higher ( < .01) compared with the medium (8.1 ± 0.4 mm Hg and 5.2 ± 0.4 mm Hg) and large AAs (7.4 ± 1.0 mm Hg and 5.4 ± 0.6 mm Hg). The net pressure gradient was lower for the case with the medium AA (4.1 ± 1.2 mm Hg) compared with the small AA (4.7 ± 0.8 mm Hg) and large AA (6.1 ± 1.4 mm Hg;  < .01).

CONCLUSIONS

We have shown that small and large AAs can increase net pressure gradient, because of the direct interaction of the Medtronic Evolut R stent with the AA (in small AA) and introducing higher level of turbulence (in large AA). AA size might need to be considered in the selection of an appropriate device for transcatheter aortic valve replacement.

摘要

目的

在本研究中,我们旨在通过评估在相同主动脉瓣环尺寸下不同升主动脉(AA)直径时瓣膜的性能和压力恢复情况,来了解美敦力Evolut R经导管主动脉瓣膜与升主动脉相互作用的作用。

方法

使用左心模拟器在生理条件下,在不同AA直径(D)的主动脉根部模型中测试26毫米的美敦力Evolut R瓣膜:小(D = 23毫米)、中(D = 28毫米)和大(D = 34毫米)。使用导管以小间隔测量瓣膜上游至下游的压力,以全面评估压力梯度和压力恢复情况。

结果

在小AA中,在收缩期测得的峰值和平均压力梯度分别为11.5±0.5毫米汞柱和7.8±0.4毫米汞柱,与中AA(8.1±0.4毫米汞柱和5.2±0.4毫米汞柱)和大AA(7.4±1.0毫米汞柱和5.4±0.6毫米汞柱)相比更高(P <.01)。中AA情况下的净压力梯度(4.1±1.2毫米汞柱)低于小AA(4.7±0.8毫米汞柱)和大AA(6.1±1.4毫米汞柱;P <.01)。

结论

我们已经表明,小AA和大AA会增加净压力梯度,这是由于美敦力Evolut R支架与AA的直接相互作用(在小AA中)以及引入更高水平的湍流(在大AA中)。在选择合适的经导管主动脉瓣置换装置时可能需要考虑AA大小。