Cao Xin, Lv Kun, Yin Xuyang, Cao Yunxi, Xu Siting, Feng Zhe, Han Yan, Tang Ye, Geng Daoying, Zhang Jun
Department of Radiology,Huashan Hospital,State Key Laboratory of Medical Neurobiology, Fudan University, Shanghai, China.
National Center for Neurological Disorders, Shanghai, China.
J Magn Reson Imaging. 2023 May;57(5):1543-1551. doi: 10.1002/jmri.28415. Epub 2022 Aug 30.
Three-dimensional (3D) contrast-enhanced T -weighted flow-sensitive black-blood (CE-T WI FSBB) is a newly developed black blood sequence by adding motion probing gradient pulses to gradient echo (GRE) sequences, which has important value for the preoperative assessment of tumor brain blood supply vessels and intratumoral microbleeds.
To compare 3D CE-T WI FSBB and 3D contrast-enhanced fast spin echo (FSE) sequence for T WI for preoperative assessment of blood vessels and microbleeds in brain tumors and to investigate the correlation between visible vessels and microbleeds.
Prospective.
One hundred and seventy-five patients with brain tumors, 65 were male, 110 were female. Including histologically confirmed 73 meningiomas, 23 schwannomas, 20 gliomas, 7 hemangioblastomas, 5 metastases, 2 lymphomas, 2 hemangiopericytomas, 2 germ cell tumors, 1 craniopharyngioma, and 1 cholesteatoma.
FIELD STRENGTH/SEQUENCE: A 3-T, CE-T WI FSBB, GRE; 3-T, CE-T WI, FSE.
Three neuroradiologists counted the number of intratumoral vessels on CE-T WI and CE-T WI FSBB images separately, and they counted the number of intratumoral microbleeds on CE-T WI FSBB images. Brain tumors were classified into grade I, grade II, and grade IV according to the World Health Organization (WHO) grading. Differences in the ability of CE-T WI FSBB and CE-T WI to display intratumoral vessels were compared. The mean counts of three observers were used to study the correlation between vessels and microbleeds.
Two-way random intraclass correlation coeficient (ICC) was used for inter-reader agreement regarding intratumoral vessel and microbleed counts, and the linear regression analysis (with F-test) was used to study the correlation between intratumoral vessels and microbleeds based on CE-T WI FSBB (α = 0.05).
Inter-reader agreements for intratumoral vessel count on CE-T WI (ICC = 0.93) and CE-T WI FSBB (ICC = 0.92), and the agreement for intratumoral microbleed count on CE-T WI FSBB (ICC = 0.99) were excellent. There were statistically significant differences in intratumoral vessel counts between CE-T WI and CE-T WI FSBB using Mann-Whitney U -test: image readers could identify more intratumoral vessels on CE-T WI FSBB images, particularly for meningiomas, schwannomas, gliomas, and WHO grade I tumors. The number of intratumoral vessels had a significant positive effect on the number of intratumoral microbleeds (microbleeds = 5.024 + 1.665 × vessels; F = 11.51).
More intratumoral vessels could potentially be identified using a 3D CE-T WI FSBB sequence compared to a CE-T WI sequence, and the number of intratumoral vessels showed a positive linear relationship with the number of intratumoral microbleeds, which might suggest that brain tumors with rich blood supply were more prone to intratumoral microbleeds.
2 TECHNICAL EFFICACY: Stage 3.
三维(3D)对比增强T加权血流敏感黑血(CE-T WI FSBB)是通过在梯度回波(GRE)序列中添加运动探测梯度脉冲而新开发的黑血序列,对肿瘤脑供血血管和瘤内微出血的术前评估具有重要价值。
比较3D CE-T WI FSBB和3D对比增强快速自旋回波(FSE)序列用于T WI对脑肿瘤血管和微出血的术前评估,并研究可见血管与微出血之间的相关性。
前瞻性研究。
175例脑肿瘤患者,男性65例,女性110例。包括组织学确诊的73例脑膜瘤、23例神经鞘瘤、20例胶质瘤、7例血管母细胞瘤、5例转移瘤、2例淋巴瘤、2例血管外皮细胞瘤、2例生殖细胞瘤、1例颅咽管瘤和1例胆脂瘤。
场强/序列:3T,CE-T WI FSBB,GRE;3T,CE-T WI,FSE。
三名神经放射科医生分别在CE-T WI和CE-T WI FSBB图像上计数瘤内血管数量,并在CE-T WI FSBB图像上计数瘤内微出血数量。根据世界卫生组织(WHO)分级将脑肿瘤分为I级、II级和IV级。比较CE-T WI FSBB和CE-T WI显示瘤内血管能力的差异。采用三名观察者的平均计数来研究血管与微出血之间的相关性。
采用双向随机组内相关系数(ICC)评估读者间对瘤内血管和微出血计数的一致性,采用线性回归分析(F检验)研究基于CE-T WI FSBB的瘤内血管与微出血之间的相关性(α = 0.05)。
读者间对CE-T WI上瘤内血管计数的一致性(ICC = 0.93)和CE-T WI FSBB上瘤内血管计数的一致性(ICC = 0.92),以及CE-T WI FSBB上瘤内微出血计数的一致性(ICC = 0.99)均极佳。使用曼-惠特尼U检验,CE-T WI和CE-T WI FSBB之间瘤内血管计数存在统计学显著差异:图像阅片者在CE-T WI FSBB图像上能识别出更多瘤内血管,尤其是对于脑膜瘤、神经鞘瘤、胶质瘤和WHO I级肿瘤。瘤内血管数量对瘤内微出血数量有显著正影响(微出血 = 5.024 + 1.665×血管;F = 11.51)。
与CE-T WI序列相比,使用3D CE-T WI FSBB序列可能识别出更多瘤内血管,且瘤内血管数量与瘤内微出血数量呈正线性关系,这可能表明血供丰富的脑肿瘤更容易发生瘤内微出血。
2级 技术效能:3级