Suppr超能文献

射波刀同步系统在呼吸相位移动肝脏肿瘤中的剂量学评估。

Dosimetric Evaluation of CyberKnife Synchrony System for Liver Tumors With Respiratory Phase Shifts.

机构信息

Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, Suita, Japan.

Department of Radiation Oncology, Osaka University Graduate School of Medicine, Suita, Japan;

出版信息

In Vivo. 2022 Nov-Dec;36(6):2861-2868. doi: 10.21873/invivo.13026.

Abstract

BACKGROUND/AIM: This study evaluated the effects of the respiratory phase shifts between liver tumor and chest wall motions on the dose distribution for the CyberKnife Synchrony respiratory tracking system (SRTS).

PATIENTS AND METHODS

Eight patients who received stereotactic body radiotherapy for hepatocellular carcinoma or liver metastases were analyzed. Three-dimensional (3D) motion of the implanted fiducial markers and vertical motion of the sternal bone were derived from the four-dimensional computed tomography (4D-CT) images acquired with a 320-row area detector CT. For each patient, Gaussian random numbers were generated for the standard deviation of the tracking error calculated from the phase shift and a literature. For each voxel of the target, the dose delivered from each beam was calculated 100 times with the random 3D offsets representing the tracking error.

RESULTS

The median respiratory phase shifts were 6.0% and 4.6% for the anterior-posterior (AP) and superior-inferior (SI) directions, respectively. The median motion tracking errors influenced by respiratory phase shifts were 1.21 mm and 0.96 mm for the AP and SI directions, respectively. The change in the dose covering 90% of the target (D) was within 1.1% when median phase shifts were considered. When evaluating the 90 percentile of the phase shifts, the D decreased up to 6.6%.

CONCLUSION

We have developed a technique to estimate the impact of the respiratory phase shifts on the dose distribution of a liver tumor treated with the SRTS. The calculation of the respiratory phase shifts from the area-detector 4D-CT will be valuable to improve the tracking accuracy of the SRTS.

摘要

背景/目的:本研究评估了肝脏肿瘤和胸壁运动之间的呼吸相位变化对 CyberKnife Synchrony 呼吸跟踪系统 (SRTS) 的剂量分布的影响。

患者和方法

分析了 8 名接受立体定向体放射治疗肝细胞癌或肝转移瘤的患者。从使用 320 排区域探测器 CT 获得的四维 CT (4D-CT) 图像中得出植入的基准标记物的三维 (3D) 运动和胸骨的垂直运动。对于每个患者,从相位偏移和文献中计算出的跟踪误差的标准偏差生成高斯随机数。对于每个靶区的每个体素,使用代表跟踪误差的随机 3D 偏移量计算 100 次从每个射束传递的剂量。

结果

前-后 (AP) 和上-下 (SI) 方向的中位呼吸相位偏移分别为 6.0%和 4.6%。受呼吸相位偏移影响的中位运动跟踪误差分别为 AP 和 SI 方向的 1.21 毫米和 0.96 毫米。当考虑中位相位偏移时,覆盖 90%靶区的剂量 (D) 的变化在 1.1%以内。当评估相位偏移的 90%分位数时,D 减少了高达 6.6%。

结论

我们已经开发了一种技术来估计呼吸相位偏移对 SRTS 治疗的肝脏肿瘤剂量分布的影响。从区域探测器 4D-CT 计算呼吸相位偏移将有助于提高 SRTS 的跟踪精度。