Suppr超能文献

采用黏附式表面电极行腮腺切除术时对面神经监测的可行性:与针电极的对比研究。

Feasibility of facial nerve monitoring using adhesive surface electrodes during parotidectomy: a comparative study with needle electrodes.

机构信息

Department of Otorhinolaryngology-Head and Neck Surgery, Pusan National University School of Medicine, Pusan National University and Biomedical Research Institute, Pusan National University Hospital, 179, Gudeok-Ro, Seo-Gu, Busan, 49241, Republic of Korea.

出版信息

Eur Arch Otorhinolaryngol. 2023 Apr;280(4):1903-1907. doi: 10.1007/s00405-022-07731-x. Epub 2022 Nov 9.

Abstract

PURPOSE

Electrophysiological monitoring of the facial nerve during parotidectomy has been reported as an adjunctive method to prevent facial nerve injury. Classically, a needle electrode is used to obtain electromyographic (EMG) signals from facial muscles during facial nerve monitoring (FNM) of parotid surgery, likewise adhesive surface electrodes. This study aimed to investigate the feasibility of performing FNM with surface electrodes during parotid surgery and to compare EMG values with needle electrodes.

METHODS

Thirty patients who underwent parotidectomy under FNM using adhesive surface and needle electrodes were included. Two pairs of adhesive surface electrodes and needle electrodes were used for FNM during parotid surgery. Mean amplitudes were collected after electrical facial nerve stimulation at 1 mA after specimen removal.

RESULTS

The mean amplitude of the adhesive surface electrodes was 226.50 ± 118.44 μV (orbicularis oculi muscle) and 469.6 ± 306.06 μV (orbicularis oris muscle), respectively. The mean amplitude of the needle electrodes was 449.85 ± 248.10 μV (orbicularis oculi muscle) and 654.66 ± 395.71 μV (orbicularis oris muscle), respectively. The mean amplitude of the orbicularis oris muscle was significantly greater than that of the orbicularis oculi. The amplitude values measured in the orbicularis oculi muscle showed significant differences between the needle and skin electrodes.

CONCLUSIONS

Facial nerve monitoring (FNM) using adhesive surface electrodes is feasible in parotid surgery. Although the mean amplitude value of the surface electrode was relatively lower than that of the needle electrode, the surface electrode is considered a feasible and safe EMG recording device for FNM in parotid surgery.

摘要

目的

在腮腺切除术期间对面神经进行电生理监测已被报道为预防面神经损伤的辅助方法。经典地,在腮腺手术的面神经监测(FNM)期间使用针电极从面部肌肉获得肌电图(EMG)信号,同样使用粘性表面电极。本研究旨在研究在腮腺手术期间使用表面电极进行 FNM 的可行性,并比较 EMG 值与针电极。

方法

纳入了 30 名在 FNM 下接受腮腺切除术的患者,使用粘性表面电极和针电极。在腮腺手术期间使用两种对粘性表面电极和针电极进行 FNM。在标本切除后以 1 mA 进行电刺激后面神经后,收集平均幅度。

结果

粘性表面电极的平均幅度分别为 226.50±118.44μV(眼轮匝肌)和 469.6±306.06μV(口轮匝肌)。针电极的平均幅度分别为 449.85±248.10μV(眼轮匝肌)和 654.66±395.71μV(口轮匝肌)。口轮匝肌的平均幅度明显大于眼轮匝肌。眼轮匝肌的幅度值在针电极和皮肤电极之间显示出显著差异。

结论

在腮腺手术中使用粘性表面电极进行面神经监测(FNM)是可行的。尽管表面电极的平均幅度值相对较低,但表面电极被认为是在腮腺手术中进行 FNM 的一种可行且安全的 EMG 记录设备。

文献AI研究员