日常 EPID 传输图像的伽马通过率与乳腺癌调强放疗计划的 PTV 覆盖相关。

Gamma passing rates of daily EPID transit images correlate to PTV coverage for breast cancer IMRT treatment plans.

机构信息

Servei de Física i Protecció Radiològica, Hospital Universitari Vall d'Hebron, Vall d'Hebron Barcelona Hospital Campus, Barcelona, Spain.

Servei d'Oncologia Radioteràpica, Hospital Universitari Vall d'Hebron, Vall d'Hebron Barcelona Hospital Campus, Barcelona, Spain.

出版信息

J Appl Clin Med Phys. 2023 May;24(5):e13913. doi: 10.1002/acm2.13913. Epub 2023 Jan 26.

Abstract

PURPOSE

The use of the transit image obtained with the electronic portal-imaging device (EPID) is becoming an extended method to perform in-vivo dosimetry. The transit images acquired during each fraction can be compared with a predicted image, if available, or with a baseline image, usually the obtained in the first fraction. This work aims to study the dosimetric impact of the failing fractions and to evaluate the appropriateness of using a baseline image in breast plans.

MATERIAL AND METHODS

Twenty breast patients treated in a Halcyon were retrospectively selected. For each patient and fraction, the treatment plan was calculated over the daily CBCT image. For each fraction, the differences respect to the treatment plan values of OARs and PTV dosimetric parameters were analyzed: ΔD , ΔD95%, ΔD98%, ΔD2%, ΔV36Gy, ΔV38.5Gy, and ΔV43.5Gy. Daily fractions were ranked according to the differences found in the dosimetric parameters between the treatment plan and the daily CBCT to establish the best fraction. The daily transit images acquired in every fraction were compared to the first fraction using the global gamma index with the Portal Dosimetry tool. The comparison was repeated using the best fraction image as a baseline. We assessed the correlation of the dosimetric differences obtained from the CBCT images-based treatment plans with the gamma index passing rates obtained using first fraction and best fraction as baseline.

RESULTS

Average values of -11.6% [-21.4%, -3.3%] and -3.2% [-1.0%, -10.3%] for the ∆PTVD98% and ∆PTVD95% per every 10% decrease in the passing rate were found, respectively. When using the best fraction as baseline patients were detected with failing fractions that were not detected with the first fraction as baseline.

CONCLUSION

The gamma passing rates of daily transit images correlate with the coverage loss parameters in breast IMRT plans. Using first fraction image as baseline can lead to the non-detectability of failing fractions.

摘要

目的

电子射野影像装置(EPID)获得的传输图像正在成为一种扩展方法,用于进行体内剂量测定。如果有预测图像可用,或者有基准图像(通常是在第一部分获得的),则可以将每次分割获得的传输图像与预测图像或基准图像进行比较。本研究旨在研究失败分割的剂量学影响,并评估在乳房计划中使用基准图像的适当性。

材料和方法

回顾性选择了 20 名在 Halcyon 接受治疗的乳腺癌患者。对于每个患者和每个部分,根据每天的 CBCT 图像计算治疗计划。对于每个部分,分析了 OAR 和 PTV 剂量学参数相对于治疗计划值的差异:ΔD、ΔD95%、ΔD98%、ΔD2%、ΔV36Gy、ΔV38.5Gy 和 ΔV43.5Gy。根据治疗计划与每日 CBCT 之间在剂量学参数上发现的差异,对每日部分进行排名,以确定最佳部分。使用 Portal Dosimetry 工具将每个部分中获得的每日传输图像与第一部分的图像进行全局伽玛指数比较。使用最佳部分图像作为基准重复比较。我们评估了基于 CBCT 图像的治疗计划中获得的剂量学差异与使用第一部分和最佳部分作为基准的伽玛指数通过率之间的相关性。

结果

当通过率每降低 10%时,分别发现∆PTVD98%和∆PTVD95%的平均值为-11.6%[-21.4%,-3.3%]和-3.2%[-1.0%,-10.3%]。当使用最佳部分作为基准时,检测到了第一部分作为基准时未检测到的失败部分。

结论

每日传输图像的伽玛通过率与乳房调强放疗计划的覆盖损失参数相关。使用第一部分图像作为基准可能导致无法检测到失败部分。