Phenylketonuria (PKU) was the first disease to be identified by the newborn screening (NBS) program. Currently, there are various methods for determining phenylalanine (Phe) values, with tandem mass spectrometry (MS/MS) being the most widely used method worldwide. We aimed to compare the MS/MS method with the fluorometric method (FM) for measuring Phe in the dried blood spot (DBS) and the efficacy of both methods in the NBS program. The FM was performed using a neonatal phenylalanine kit and a VICTOR2 D fluorometer. The MS/MS method was performed using a NeoBase 2 kit and a Waters Xevo TQD mass spectrometer. The Phe values measured with the MS/MS method were compared to those determined by the FM. The cut-off value for the NBS program was set at 120 µmol/L for FM and 85 µmol/L for MS/MS. We analyzed 54,934 DBS. The measured Phe values varied from 12 to 664 µmol/L, with a median of 46 µmol/L for the MS/MS method and from 10 to 710 µmol/L, with a median of 70 µmol/L for the FM. The Bland-Altman analysis indicated a bias of -38.9% (-23.61 µmol/L) with an SD of 21.3% (13.89 µmol/L) when comparing the MS/MS method to the FM. The Phe value exceeded the cut-off in 187 samples measured with FM and 112 samples measured with MS/MS. The FM had 181 false positives, while the MS/MS method had 106 false positives. Our study showed that the MS/MS method gives lower results compared to the FM. Despite that, none of the true positives would be missed, and the number of false-positive results would be significantly lower compared to the FM.
苯丙酮尿症(PKU)是新生儿筛查(NBS)项目中最早确定的疾病之一。目前,有多种方法可用于测定苯丙氨酸(Phe)值,串联质谱法(MS/MS)是世界范围内最广泛使用的方法。我们旨在比较 MS/MS 法与荧光法(FM)测定干血斑(DBS)中 Phe 的效果,并比较两种方法在 NBS 项目中的效果。FM 使用新生儿苯丙氨酸试剂盒和 VICTOR2 D 荧光计进行。MS/MS 方法使用 NeoBase 2 试剂盒和 Waters Xevo TQD 质谱仪进行。将 MS/MS 方法测量的 Phe 值与 FM 测定的值进行比较。NBS 程序的截止值设定为 FM 为 120 μmol/L,MS/MS 为 85 μmol/L。我们分析了 54934 个 DBS。测定的 Phe 值范围为 12-664 μmol/L,MS/MS 方法的中位数为 46 μmol/L,FM 的范围为 10-710 μmol/L,中位数为 70 μmol/L。Bland-Altman 分析表明,当将 MS/MS 方法与 FM 进行比较时,存在 -38.9%(-23.61 μmol/L)的偏倚,SD 为 21.3%(13.89 μmol/L)。用 FM 测量的 187 个样本和用 MS/MS 测量的 112 个样本的 Phe 值超过了截止值。FM 有 181 个假阳性,而 MS/MS 方法有 106 个假阳性。我们的研究表明,MS/MS 方法与 FM 相比,结果较低。尽管如此,不会错过任何真正的阳性结果,而且与 FM 相比,假阳性结果的数量会显著降低。