心脏植入式电子设备与连续血流左心室辅助设备之间的设备-设备相互作用。

Device-device interaction between cardiac implantable electronic devices and continuous-flow left ventricular assist devices.

机构信息

Division of Cardiovascular Medicine, Indiana University, Indianapolis, Indiana.

Division of Cardiovascular Medicine, Indiana University, Indianapolis, Indiana.

出版信息

Heart Rhythm. 2023 Jun;20(6):918-926. doi: 10.1016/j.hrthm.2023.02.020. Epub 2023 Feb 28.

Abstract

The current design of an innovative left ventricular assist device (LVAD) makes use of magnetic levitation technology, which enables the rotors of the device to be completely suspended by magnetic force, reducing friction and blood or plasma damage. However, this electromagnetic field can result in electromagnetic interference (EMI), which can interfere with proper functioning of another cardiac implantable electronic device (CIED) in its direct proximity. Approximately 80% of patients with an LVAD have a CIED, most frequently an implantable cardioverter-defibrillator (ICD). Several device-device interactions have been reported, including EMI-induced inappropriate shocks, inability to establish telemetry connection, EMI-induced premature battery depletion, undersensing by the device, and other CIED malfunctions. Unfortunately, additional procedures, including generator exchange, lead adjustment, and system extraction, are frequently required because of these interactions. In some circumstances, the additional procedure might be preventable or avoidable with appropriate solutions. In this article, we describe how EMI from the LVAD impacts the functionality of the CIED and provide possible management options, including manufacturer-specific information, for the current CIEDs (eg, transvenous and leadless pacemakers, transvenous and subcutaneous ICDs, and transvenous cardiac resynchronization therapy pacemakers and ICDs).

摘要

目前,一种创新的左心室辅助装置(LVAD)的设计采用了磁悬浮技术,使装置的转子完全通过磁力悬浮,从而减少摩擦和血液或血浆损伤。然而,这个电磁场可能会导致电磁干扰(EMI),从而干扰其附近的另一个心脏植入式电子设备(CIED)的正常运行。大约 80%的 LVAD 患者都有 CIED,最常见的是植入式心律转复除颤器(ICD)。已经报告了几种设备-设备相互作用,包括 EMI 引起的不适当电击、无法建立遥测连接、EMI 引起的电池过早耗尽、设备检测不足以及其他 CIED 故障。不幸的是,由于这些相互作用,经常需要进行额外的程序,包括发电机更换、导联调整和系统提取。在某些情况下,通过适当的解决方案,额外的程序可能是可以预防或避免的。在本文中,我们描述了 LVAD 的 EMI 如何影响 CIED 的功能,并提供了当前 CIED(例如,经静脉和无导线起搏器、经静脉和皮下 ICD 以及经静脉心脏再同步治疗起搏器和 ICD)的可能的管理选择,包括制造商特定的信息。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验