BK21 FOUR Team and Integrated Research Institute for Drug Development, College of Pharmacy, Dongguk University, Gyeonggi 10326, Republic of Korea.
College of Pharmacy, Mokpo National University, Jeonnam 58554, Republic of Korea.
Int J Pharm. 2023 Jul 25;642:123091. doi: 10.1016/j.ijpharm.2023.123091. Epub 2023 Jun 1.
Subvisible particles generated during the preparation or administration of biopharmaceuticals might increase the risk of immunogenicity, inflammation, or organ dysfunction. To investigate the impact of an infusion system on the level of subvisible particles, we compared two types of infusion sets based on peristaltic movement (Medifusion DI-2000 pump) and a gravity-based infusion system (Accu-Drip) using intravenous immunoglobulin (IVIG) as a model drug. The peristaltic pump was found to be more susceptible to particle generation compared to the gravity infusion set owing to the stress generated due to constant peristaltic motion. Moreover, the 5-µm in-line filter integrated into the tubing of the gravity-based infusion set further contributed to the reduction of particles mostly in the range ≥ 10 µm. Furthermore, the filter was also able to maintain the particle level even after the pre-exposure of samples to silicone oil-lubricated syringes, drop shock, or agitation. Overall, this study suggests the need for the selection of an appropriate infusion set equipped with an in-line filter based on the sensitivity of the product.
在生物制药的制备或给药过程中产生的亚可见颗粒可能会增加免疫原性、炎症或器官功能障碍的风险。为了研究输液系统对亚可见颗粒水平的影响,我们使用静脉注射免疫球蛋白(IVIG)作为模型药物,比较了两种基于蠕动运动的输液套件(Medifusion DI-2000 泵)和基于重力的输液系统(Accu-Drip)。与重力输液套件相比,蠕动泵更容易产生颗粒,这是由于由于持续的蠕动运动产生的应力。此外,集成在重力输液套件的管内的 5μm 在线过滤器进一步有助于减少主要在 10μm 以上范围内的颗粒。此外,即使在样品预先暴露于硅基油润滑注射器、滴注冲击或搅拌后,该过滤器仍能够维持颗粒水平。总的来说,这项研究表明需要根据产品的敏感性选择配备在线过滤器的合适输液套件。