Diabetic Retinopathy Screening Using Non-Mydriatic Fundus Camera in Primary Health Care Settings - A Multicenter Study from Saudi Arabia.

作者信息

Alabdulwahhab Khalid Mohammad

机构信息

Department of Ophthalmology, College of Medicine, Majmaah University, Majmaah, Kingdom of Saudi Arabia.

出版信息

Int J Gen Med. 2023 Jun 5;16:2255-2262. doi: 10.2147/IJGM.S410197. eCollection 2023.

Abstract

BACKGROUND

Screening of diabetic retinopathy (DR) using the current digital imaging facilities in a primary health care setting is still in its early stages in Saudi Arabia. This study aims to reduce the risk of vision impairment and blindness among known diabetic people through early identification by general practitioners (GP) in a primary health care setting in Saudi Arabia. The objective of this study was to evaluate the accuracy of diabetic retinopathy (DR) detection by general practitioners (GPs) by comparing the agreement of DR assessment between GPs and ophthalmologists' assessment as a gold standard.

METHODS

A hospital-based, six-month cross-sectional study was conducted, and the participants were type 2 diabetic adults from the diabetic registries of seven rural PHCs, in Saudi Arabia. After medical examination, the participants were then evaluated by fundus photography using a non-mydriatic fundus camera without medication for mydriasis. Presence or absence of DR was graded by the trained GPs in the PHCs and then compared with the grading of an ophthalmologist which was taken as a reference or a gold standard.

RESULTS

A total of 899 diabetic patients were included, and the mean age of the patients was 64.89 ± 11.01 years. The evaluation by the GPs had a sensitivity of 80.69 [95% CI 74.8-85.4]; specificity of 92.23 [88.7-96.3]; positive predictive value, 74.1 [70.4-77.0]; negative predictive value, 73.34 [70.6-77.9]; and an accuracy of 84.57 [81.8-89.88]. For the consensus of agreement the adjusted kappa coefficient was from 0.74 to 0.92 for the DR.

CONCLUSION

This study demonstrates that trained GPs in rural health centers are able to provide reliable detection results of DR from fundus photographs. The study highlights the need for early DR screening programs in the rural areas of Saudi Arabia to facilitate early identification of the condition and to lessen impact of blindness due to diabetes.

摘要

背景

在沙特阿拉伯,利用初级卫生保健机构现有的数字成像设备筛查糖尿病视网膜病变(DR)仍处于早期阶段。本研究旨在通过沙特阿拉伯初级卫生保健机构的全科医生(GP)进行早期识别,降低已知糖尿病患者视力损害和失明的风险。本研究的目的是通过比较全科医生与眼科医生作为金标准的DR评估一致性,评估全科医生检测糖尿病视网膜病变(DR)的准确性。

方法

进行了一项基于医院的为期六个月的横断面研究,参与者为沙特阿拉伯七个农村初级卫生保健中心糖尿病登记处的2型糖尿病成年人。经过医学检查后,使用非散瞳眼底相机在不使用散瞳药物的情况下对参与者进行眼底摄影评估。初级卫生保健中心经过培训的全科医生对DR的存在与否进行分级,然后与作为参考或金标准的眼科医生的分级进行比较。

结果

共纳入899例糖尿病患者,患者平均年龄为64.89±11.01岁。全科医生的评估敏感性为80.69[95%CI 74.8-85.4];特异性为92.23[88.7-96.3];阳性预测值为74.1[70.4-77.0];阴性预测值为73.34[70.6-77.9];准确性为84.57[81.8-89.88]。对于一致性共识,DR的调整kappa系数为0.74至0.92。

结论

本研究表明,农村卫生中心经过培训的全科医生能够从眼底照片中提供可靠的DR检测结果。该研究强调了沙特阿拉伯农村地区需要早期DR筛查项目,以便于早期识别病情并减轻糖尿病导致失明的影响。