Adolfi Dario, Grangeiro Manassés Tercio Vieira, Ausiello Pietro, Bottino Marco Antonio, Tribst João Paulo Mendes
Department of Dental Materials and Prosthodontics, Institute of Science and Technology, São Paulo State University (UNESP), São José Dos Campos 12220-000, Brazil.
School of Dentistry, University of Naples Federico II, Via S. Pansini 5, 80131 Naples, Italy.
Materials (Basel). 2023 Jul 6;16(13):4848. doi: 10.3390/ma16134848.
This study investigated the effects of antirotational titanium bases on the mechanical behavior of CAD/CAM titanium bases used for implant-supported prostheses. The aim was to assess the impact on the marginal fit, fatigue behavior, stress concentration, and fracture load of implant-supported CAD/CAM zirconia crowns. Forty titanium implants were divided into two groups: those with antirotational titanium bases (ARs) and those with rotational titanium bases (RTs). Torque loosening and vertical misfit were evaluated before and after cyclic fatigue testing (200 N, 2 Hz, 2 × 10 cycles). Fracture resistance was assessed using a universal testing machine (1 mm/min, 1000 kgf), and failed specimens were examined with microscopy. Three-dimensional models were created, and FEA was used to calculate stress. Statistical analysis was performed on the in vitro test data using two-way analysis of variance and Tukey's test (α = 0.5). Results show that the presence of an antirotational feature between the implant and titanium base reduced preload loss and stress concentration compared to rotational titanium bases. However, there were no differences in vertical misfit and resistance to compressive load.
本研究调查了抗旋转钛基托对用于种植体支持修复体的CAD/CAM钛基托力学行为的影响。目的是评估其对种植体支持的CAD/CAM氧化锆冠边缘适合性、疲劳行为、应力集中和断裂载荷的影响。40颗钛种植体被分为两组:一组使用抗旋转钛基托(ARs),另一组使用可旋转钛基托(RTs)。在循环疲劳测试(200 N,2 Hz,2×10⁶ 次循环)前后评估扭矩松动和垂直不贴合情况。使用万能试验机(1 mm/min,1000 kgf)评估抗折性,对失效样本进行显微镜检查。创建三维模型,并使用有限元分析计算应力。使用双向方差分析和Tukey检验(α = 0.05)对体外测试数据进行统计分析。结果表明,与可旋转钛基托相比,种植体与钛基托之间抗旋转特征的存在减少了预紧力损失和应力集中。然而,在垂直不贴合和抗压载荷抗性方面没有差异。