Kido Takahisa, Ono Tomohiro, Nakamura Mitsuhiro, Ishihara Yoshitomo, Itoh Hiroyuki, Matsugi Kiyotomo, Yoshimoto Atsushi, Kishigami Yukako, Mizowaki Takashi
Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
Department of Radiation Oncology and Image-Applied Therapy, Kyoto University, Kyoto, Japan.
Phys Med. 2023 Aug;112:102645. doi: 10.1016/j.ejmp.2023.102645. Epub 2023 Jul 19.
Single-isocenter stereotactic radiotherapy for multiple brain metastases requires highly accurate treatment delivery at off-isocenter positions (off-iso). This study aimed to verify the beam-positioning errors at off-iso using a newly developed phantom tested at multiple institutions.
The off-iso phantom comprised five stainless-steel balls with a 3-mm diameter placed at the center and at four peripheral positions on a diagonal line. Each ball was placed 3.5 cm apart along each of the three axes. Two patterns of the phantom setup were defined as 0° and 90° phantom rotations to evaluate the beam-positioning error, which is the distance between the center of the ball and the irradiated field on the electronic portal imaging device. Furthermore, the reproducibility of the beam-positioning errors was verified by evaluating their standard deviation (SD) at a single institution, which included five measurements for two treatment machines. The errors were evaluated at multiple institutions using eight treatment machines.
The measurement time from setup to image acquisition was approximately 20 min for two patterns. The SD of the beam-positioning errors in the reproducibility tests was 0.41 mm. In the multi-institutional evaluation, the beam-positioning error at the isocenter position was within 1.00 mm of the AAPM-RSS tolerance, with the exception of two linacs. The largest beam-positioning error (1.36 mm) was observed 7.5 cm away from the isocenter in three directions at a gantry angle of 180°.
The developed phantom can be applied as a new tool for establishing beam-positioning errors in single-isocenter stereotactic radiotherapy at off-isocenter positions.
针对多发脑转移瘤的单等中心立体定向放射治疗需要在等中心外位置(非等中心)进行高度精确的治疗投照。本研究旨在使用一种新开发的模体,在多个机构进行测试,以验证非等中心处的射束定位误差。
非等中心模体由五个直径3毫米的不锈钢球组成,放置在中心以及对角线上的四个周边位置。每个球沿三个轴中的每一个轴相距3.5厘米放置。模体设置的两种模式定义为0°和90°模体旋转,以评估射束定位误差,即球心与电子射野影像装置上照射野之间的距离。此外,通过在单个机构评估其标准差(SD)来验证射束定位误差的可重复性,其中包括对两台治疗机器进行五次测量。使用八台治疗机器在多个机构评估误差。
两种模式从设置到图像采集的测量时间约为20分钟。可重复性测试中射束定位误差的标准差为0.41毫米。在多机构评估中,除两台直线加速器外,等中心位置的射束定位误差在AAPM-RSS容差的1.00毫米范围内。在机架角度为180°时,在三个方向上距离等中心7.5厘米处观察到最大射束定位误差(1.36毫米)。
所开发的模体可作为一种新工具,用于确定单等中心立体定向放射治疗在非等中心位置的射束定位误差。