Suppr超能文献

用于前列腺超声检查的标准经直肠探头与小型经直肠探头的对比研究。

Comparative study of standard and small transrectal transducers for prostate ultrasonography.

作者信息

Hwang Sung Il, Ahn Hyungwoo, Lee Hak Jong, Jung Sung Il

机构信息

Department of Radiology, Seoul National University Bundang Hospital, Seongnam, Korea.

Department of Radiology, College of Medicine, Seoul National University, Seoul, Korea.

出版信息

Ultrasonography. 2023 Oct;42(4):555-560. doi: 10.14366/usg.23084. Epub 2023 Jun 30.

Abstract

PURPOSE

The objective of this study was to evaluate pain and image quality associated with the use of two different ultrasound transducers.

METHODS

Fifty healthy male participants aged 30 years or older were prospectively enrolled. All ultrasound procedures were performed using a V8 machine (Samsung Medison, Seoul, Korea) equipped with EA2-11 (conventional) and miniER7 (small-caliber) transrectal transducers, operated by a single genitourinary radiologist. To minimize bias, one group of volunteers underwent ultrasonography with the conventional transducer first, followed by the small transducer. For the remaining participants, the examinations were performed in the opposite order. Ultrasonography, including the measurement of total prostate and transitional zone volumes, was conducted in accordance with standard practice. After testing with both probes, participants were asked to rate their pain on a 10-point numerical rating scale (NRS). A radiologist then evaluated the quality of the images acquired with each probe using a 5-point numeric scale and compared the prostate volume measurements obtained by each method.

RESULTS

The mean NRS scores associated with the conventional and small transducers were 4.7±1.8 and 2.7±1.2, respectively (P<0.05). The mean ultrasound image qualities from the two transducers were statistically similar (4.78 and 4.74, P>0.05). The whole prostate gland volume as measured with the conventional transducer (mean±standard deviation, 24.2±9.1 mL) was greater than the measurement (22.1±8.7 mL) obtained with the small-caliber transducer (P<0.05). However, only two of the 50 whole gland volume measurements differed by more than two standard deviations.

CONCLUSION

The use of a small transrectal probe significantly reduced pain without compromising image quality.

摘要

目的

本研究的目的是评估使用两种不同超声换能器时的疼痛情况和图像质量。

方法

前瞻性纳入了50名30岁及以上的健康男性参与者。所有超声检查均使用配备EA2 - 11(传统型)和miniER7(小口径)经直肠换能器的V8机器(三星麦迪逊,韩国首尔)进行,由一名泌尿生殖放射科医生操作。为尽量减少偏差,一组志愿者先使用传统换能器进行超声检查,然后再使用小换能器。其余参与者则按相反顺序进行检查。超声检查包括按照标准操作测量前列腺总体积和移行区体积。在使用两种探头进行检查后,要求参与者在10分数字评分量表(NRS)上对他们的疼痛程度进行评分。然后,一名放射科医生使用5分数字量表评估每个探头获取的图像质量,并比较每种方法获得的前列腺体积测量值。

结果

与传统换能器和小换能器相关的平均NRS评分分别为4.7±1.8和2.7±1.2(P<0.05)。两种换能器的平均超声图像质量在统计学上相似(分别为4.78和4.74,P>0.05)。用传统换能器测量的整个前列腺体积(平均值±标准差,24.2±9.1 mL)大于用小口径换能器测量的值(22.1±8.7 mL)(P<0.05)。然而,在50个全腺体积测量值中,只有两个相差超过两个标准差。

结论

使用小口径经直肠探头可显著减轻疼痛,且不影响图像质量。