A Digital Box and Block Test for Hand Dexterity Measurement: Instrument Validation Study.

作者信息

Prochaska Eveline, Ammenwerth Elske

机构信息

Institute of Applied Electronics and Technical Informatics, University of Applied Sciences Campus Vienna, Vienna, Austria.

Institute of Medical Informatics, Private University for Health Sciences and Health Technology, Hall in Tirol, Austria.

出版信息

JMIR Rehabil Assist Technol. 2023 Sep 15;10:e50474. doi: 10.2196/50474.

Abstract

BACKGROUND

The Box and Block Test (BBT) measures unilateral gross manual dexterity and is widely used in clinical settings with a wide range of populations, including older people and clients with neurological disorders.

OBJECTIVE

In this study, we present a newly developed digitized version of the BBT, called the digital BBT (dBBT). The physical design is similar to the original BBT, but the dBBT contains digital electronics that automate the test procedure, timing, and score measurement. The aim of this study is to investigate the validity and reliability of the dBBT.

METHODS

We performed measurements at 2 time points for 29 healthy participants. BBT and dBBT were used at the first measurement time point, and dBBT was used again at the second measurement time point. Concurrent validity was assessed using the correlation between BBT and dBBT, the paired t test, and the Bland-Altman analysis. Test-retest reliability and interrater reliability were examined using the interclass correlation coefficient (ICC) by repeated measures with the dBBT within an interval of 10 days.

RESULTS

Our results showed moderate concurrent validity (r=0.48, P=.008), moderate test-retest reliability (ICC 0.72, P<.001), a standard error of measurement of 3.1 blocks, and the smallest detectable change at a 95% CI of 8.5 blocks. Interrater reliability was moderate with an ICC of 0.67 (P=.02). The Bland-Altman analysis showed sufficient accuracy of the dBBT in comparison with the conventional BBT.

CONCLUSIONS

The dBBT can contribute to objectifying the measurement of gross hand dexterity without losing its important characteristics and is simple to implement.

摘要

背景

箱块测试(BBT)用于测量单侧手部总体灵巧性,广泛应用于包括老年人和神经障碍患者在内的各类人群的临床环境中。

目的

在本研究中,我们展示了一种新开发的BBT数字化版本,称为数字BBT(dBBT)。其物理设计与原始BBT相似,但dBBT包含数字电子设备,可自动执行测试程序、计时和分数测量。本研究的目的是调查dBBT的有效性和可靠性。

方法

我们对29名健康参与者在两个时间点进行了测量。在第一个测量时间点使用BBT和dBBT,在第二个测量时间点再次使用dBBT。通过BBT与dBBT之间的相关性、配对t检验和布兰德-奥特曼分析评估同时效度。使用组内相关系数(ICC),在10天的间隔内对dBBT进行重复测量,以检验重测可靠性和评分者间可靠性。

结果

我们的结果显示出中等程度的同时效度(r = 0.48,P = 0.008)、中等程度的重测可靠性(ICC 0.72,P < 0.001)、测量标准误差为3.1个方块,以及95%置信区间下最小可检测变化为8.5个方块。评分者间可靠性中等,ICC为0.67(P = 0.02)。布兰德-奥特曼分析表明,与传统BBT相比,dBBT具有足够的准确性。

结论

dBBT有助于在不失其重要特征的情况下客观化总体手部灵巧性的测量,且易于实施。