This study aims to evaluate the reliability and validity of using MyotonPRO to quantify the mechanical properties of the muscle-tendon unit through in vivo measurements and preliminary in situ measurements using formalin-fixed tissues.
The mechanical properties of gastrocnemii and the Achilles tendon of 12 healthy adults (six males and six females, 34.9 ± 5.8 years) were examined for in vivo test twice within a day and once post-24 hours using MyotonPRO, while nine human cadavers (formalin-fixed, 3 males and 6 females, 89.9 ± 5.1 years) were assessed for preliminary in situ test with identical time schedule to evaluate the within-day and inter-day reliability and validity.
In vivo tests had very high within-day (ICC: 0.96-0.99) and inter-day reliability (ICC: 0.83-0.96), while in situ tests (formalin-fixed tissues) showed high within-day (ICC: 0.87-0.99) and inter-day reliability (ICC: 0.76-0.98) for the results of tone and stiffness. There was no significant difference in the stiffness of the free part of the Achilles tendon between in vivo and in situ conditions. The stiffness of the lateral gastrocnemius ( = 0.55, = 0.018), proximal part of the Achilles tendon ( = 0.56, = 0.015), and free part of the Achilles tendon ( = 0.47, = 0.048) before removing the skin was significantly correlated with that after removing the skin condition.
The findings of the current study suggest that MyotonPRO is reliable and valid for evaluating tendon stiffness both in vivo and in situ (formalin-fixed tissues).
本研究旨在评估 MyotonPRO 通过体内测量和使用福尔马林固定组织的初步原位测量来量化肌肉-肌腱单元机械特性的可靠性和有效性。
使用 MyotonPRO 分别在一天内两次和 24 小时后对 12 名健康成年人(6 名男性和 6 名女性,34.9±5.8 岁)的腓肠肌和跟腱进行了体内测试,同时对 9 具人体尸体(福尔马林固定,3 名男性和 6 名女性,89.9±5.1 岁)进行了初步原位测试,评估了日内和日间的可靠性和有效性。
体内测试具有非常高的日内(ICC:0.96-0.99)和日间可靠性(ICC:0.83-0.96),而原位测试(福尔马林固定组织)显示出高日内(ICC:0.87-0.99)和日间可靠性(ICC:0.76-0.98)的结果。跟腱自由部分的硬度在体内和原位条件下没有显著差异。外侧腓肠肌( = 0.55, = 0.018)、跟腱近端( = 0.56, = 0.015)和跟腱自由部分( = 0.47, = 0.048)在去除皮肤之前的硬度与去除皮肤后的硬度显著相关。
本研究结果表明,MyotonPRO 可用于评估体内和原位(福尔马林固定组织)的肌腱硬度,具有可靠性和有效性。