J Acoust Soc Am. 2023 Nov 1;154(5):2858-2868. doi: 10.1121/10.0022324.
Ultrasonic backscatter techniques are being developed to detect changes in bone caused by osteoporosis and other diseases. Backscatter measurements performed at peripheral skeletal sites such as the heel may place the interrogated region of bone tissue in the acoustic near field of the transducer. The purpose of this study is to investigate how measurements in the near field affect backscatter parameters used for ultrasonic bone assessment. Ultrasonic measurements were performed in a water tank using a planar 2.25 MHz transducer. Signals were acquired for five transducer-specimen distances: N/4, N/2, 3 N/4, N, and 5 N/4, where N is the near-field distance, a location that represents the transition from the near field to far field. Five backscatter parameters previously identified as potentially useful for ultrasonic bone assessment purposes were measured: apparent integrated backscatter, frequency slope of apparent backscatter (FSAB), frequency intercept of apparent backscatter, normalized mean of the backscatter difference, and backscatter amplitude decay constant. All five parameters depended on transducer-specimen distance to varying degrees with FSAB exhibiting the greatest dependence on distance. These results suggest that laboratory studies of bone should evaluate the performance of backscatter parameters using transducer-specimen distances that may be encountered clinically including distances where the ultrasonically interrogated region is in the near field of the transducer.
超声背散射技术正被开发用于检测骨质疏松症和其他疾病引起的骨骼变化。在脚跟等外周骨骼部位进行的背散射测量可能会使骨组织的被探测区域处于换能器的近场中。本研究旨在探究近场中的测量如何影响用于超声骨评估的背散射参数。使用平面 2.25 MHz 换能器在水箱中进行了超声测量。针对五种换能器-样本距离采集了信号:N/4、N/2、3N/4、N 和 5N/4,其中 N 是近场距离,代表从近场到远场的过渡位置。测量了五个先前被确定为对超声骨评估有潜在用处的背散射参数:表观积分背散射、表观背散射频率斜率 (FSAB)、表观背散射频率截距、背散射差值的归一化均值,以及背散射幅度衰减常数。所有五个参数都在不同程度上取决于换能器-样本距离,其中 FSAB 对距离的依赖性最大。这些结果表明,骨的实验室研究应使用可能在临床中遇到的换能器-样本距离来评估背散射参数的性能,包括超声探测区域处于换能器近场的距离。