Abstract

AIM

In 2017, the Leiden University Medical Centre implemented a secondary alarm system using handheld devices to ensure accurate patient monitoring on the single room NICU. Initially, alarms remained active on the handheld devices until one of the caregivers in the alarm chain accepted the alarm. In 2020, a bidirectional communication protocol (BCP) was implemented, enabling automated withdrawal of resolved alarms. The aim of this study was to evaluate the effect of this implementation on the alarm duration and pressure.

METHODS

Data of all alarms of the secondary alarm chain in the 90 days before and after the implementation were analysed and compared between both periods.

RESULTS

Following the implementation of the BCP, 60% of the alarms were withdrawn before the designated nurse responded. Despite a significant higher total number of alarms, the median alarm duration decreased from 9 (7-14) to 6 (4-10) s, the acceptance rate of the designated nurse increased from 93% to 95% and the median time of alarm sounding per phone per hour significantly decreased from 71 (51-101) to 51 (35-69) s following implementation of the BCP.

CONCLUSION

This study showed that automated withdrawal of resolved alarms significantly reduces alarm duration and pressure on a NICU.

摘要

目的

2017 年,莱顿大学医学中心在单人间新生儿重症监护病房(NICU)中使用手持设备实施了二级报警系统,以确保对患者进行准确监测。最初,警报会在手持设备上保持活动状态,直到报警链中的一名护理人员接受警报。2020 年,实施了双向通信协议(BCP),实现了已解决警报的自动撤防。本研究旨在评估这一实施对报警持续时间和压力的影响。

方法

分析并比较了实施 BCP 前后 90 天内二级报警链的所有警报数据,并在这两个时间段之间进行比较。

结果

实施 BCP 后,有 60%的警报在指定护士响应之前被撤防。尽管总报警数显著增加,但中位数报警持续时间从 9(7-14)秒降至 6(4-10)秒,指定护士的接受率从 93%提高到 95%,每小时每部电话的报警声中位数时间从 71(51-101)秒显著降至 51(35-69)秒。

结论

本研究表明,自动撤防已解决的警报可显著降低 NICU 的报警持续时间和压力。