评估视网膜病变中光学相干断层扫描外视网膜带长度的临床相关性和可靠性:一项横断面研究。

Evaluating the clinical relevance and reliability of outer retinal band length on optical coherence tomography in retinal disease: a cross-sectional study.

机构信息

School of Optometry and Vision Science, University of New South Wales, Sydney, New South Wales, Australia.

Centre for Eye Health, University of New South Wales, Sydney, New South Wales, Australia.

出版信息

BMJ Open. 2023 Dec 12;13(12):e077874. doi: 10.1136/bmjopen-2023-077874.

Abstract

OBJECTIVES

Hyper-reflective outer retinal band (HORB) disruptions are reported across a range of retinal disease, yet a reliable, easily implemented assessment method and thorough evaluation of their association to retinal disease is lacking. The purpose of the study was to assess the reliability of using magnitude estimation to evaluate HORB length and determine its association to visual acuity and retinal disease.

DESIGN

Cross-sectional, retrospective study.

SETTING

Patients attending a secondary eye care clinic in Sydney, Australia.

PARTICIPANTS

2039 unique consecutive patients were screened for inclusion between 2 November and 18 January 2021, and 600 were included in the study population. Patients were included if they were referred from primary care, presented for an initial, comprehensive eye examination during the study period, imaged with optical coherence tomography during their visit and over 18 years of age.

PRIMARY OUTCOME

Reliability of HORB length estimations and associations to clinical outcomes.

RESULTS

Intragrader (intraclass correlation coefficient, ICC=0.81; ICC=0.91) and intergrader (ICC=0.78-0.79; ICC=0.75-0.88) agreement of HORB length was good to excellent. HORB length was significantly associated with age (p<0.001, β=-0.22 to -0.24) and refractive error (p<0.001, β=0.12-0.16) at all B-scan locations. Visual acuity (p=0.001, β=-0.13) was associated with the primary outcome for foveal B-scans and eccentricity (p=0.002, β=-0.13) and device type (p=0.002, β=0.13) for non-foveal B-scans. Glaucoma was associated with HORB length on univariate analysis (p=0.05-0.06, β=-0.08); however, multivariate analysis revealed no significant association between HORB length and retinal disease.

CONCLUSION

HORB length is reliably assessed using magnitude estimation and may be useful as a surrogate biomarker of visual acuity. Several factors affect HORB length estimations, which may contribute to the lack of association to retinal disease and highlights the need for covariable adjustment when examining HORB disruptions.

摘要

目的

在外层视网膜高反射带(HORB)在一系列视网膜疾病中均有报道,但缺乏可靠且易于实施的评估方法和对其与视网膜疾病相关性的全面评估。本研究旨在评估使用量级估计评估 HORB 长度的可靠性,并确定其与视力和视网膜疾病的相关性。

设计

横断面、回顾性研究。

地点

澳大利亚悉尼的二级眼科诊所。

参与者

2039 名患者接受了 2021 年 11 月 2 日至 1 月 18 日的纳入筛选,其中 600 名纳入研究人群。如果患者从初级保健转诊,在研究期间接受初次全面眼科检查,就诊期间接受光学相干断层扫描成像,且年龄超过 18 岁,则将其纳入研究。

主要结局

HORB 长度评估的可靠性和与临床结局的相关性。

结果

HORB 长度的组内(组内相关系数,ICC=0.81;ICC=0.91)和组间(ICC=0.78-0.79;ICC=0.75-0.88)一致性良好到优秀。HORB 长度与年龄(p<0.001,β=-0.22 至 -0.24)和屈光不正(p<0.001,β=0.12-0.16)在所有 B 扫描位置均显著相关。视力(p=0.001,β=-0.13)与中心凹 B 扫描的主要结局相关,而离焦(p=0.002,β=-0.13)和设备类型(p=0.002,β=0.13)与非中心凹 B 扫描相关。青光眼在单变量分析中与 HORB 长度相关(p=0.05-0.06,β=-0.08);然而,多变量分析显示 HORB 长度与视网膜疾病之间无显著相关性。

结论

使用量级估计可靠地评估 HORB 长度,并且可能作为视力的替代生物标志物有用。几个因素会影响 HORB 长度的估计,这可能是导致其与视网膜疾病缺乏相关性的原因,并强调在检查 HORB 中断时需要进行协变量调整。