Department of Anesthesiology, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074 Aachen, Germany.
Informatics 11-Embedded Software, RWTH Aachen University, 52074 Aachen, Germany.
Biosensors (Basel). 2023 Dec 4;13(12):1012. doi: 10.3390/bios13121012.
The occurrence of thrombus formation within an extracorporeal membrane oxygenator is a common complication during extracorporeal membrane oxygenation therapy and can rapidly result in a life-threatening situation due to arterial thromboembolism, causing stroke, pulmonary embolism, and limb ischemia in the patient. The standard clinical practice is to monitor the pressure at the inlet and outlet of oxygenators, indicating fulminant, obstructive clot formation indicated by an increasing pressure difference (ΔP). However, smaller blood clots at early stages are not detectable. Therefore, there is an unmet need for sensors that can detect blood clots at an early stage to minimize the associated thromboembolic risks for patients. This study aimed to evaluate if forward scattered light (FSL) measurements can be used for early blood clot detection and if it is superior to the current clinical gold standard (pressure measurements). A miniaturized in vitro test circuit, including a custom-made test chamber, was used. Heparinized human whole blood was circulated through the test circuit until clot formation occurred. Four LEDs and four photodiodes were placed along the sidewall of the test chamber in different positions for FSL measurements. The pressure monitor was connected to the inlet and the outlet to detect changes in ΔP across the test chamber. Despite several modifications in the LED positions on the test chamber, the FSL measurements could not reliably detect a blood clot within the in vitro test circuit, although the pressure measurements used as the current clinical gold standard detected fulminant clot formation in 11 independent experiments.
体外膜肺氧合器(ECMO)中血栓形成的发生是 ECMO 治疗过程中的常见并发症,由于动脉血栓栓塞,可迅速导致危及生命的情况,导致患者中风、肺栓塞和肢体缺血。标准的临床实践是监测氧合器的进出口压力,表明由于压力差(ΔP)增加而导致暴发性、阻塞性血栓形成。然而,早期较小的血栓无法检测到。因此,需要能够早期检测到血栓的传感器,以最大程度地降低患者相关的血栓栓塞风险。本研究旨在评估前向散射光(FSL)测量是否可用于早期检测血栓,以及其是否优于当前的临床金标准(压力测量)。使用了一个小型化的体外测试电路,包括一个定制的测试室。肝素化的人体全血在测试电路中循环,直到形成血栓。四个 LED 和四个光电二极管放置在测试室的侧壁上的不同位置,用于 FSL 测量。压力监视器连接到入口和出口,以检测测试室中 ΔP 的变化。尽管在测试室上对 LED 位置进行了几次修改,但 FSL 测量无法可靠地检测到体外测试电路中的血栓,尽管用作当前临床金标准的压力测量在 11 个独立实验中检测到了暴发性血栓形成。