A novel low-power pure-cut hot snare polypectomy for 10-14 mm colorectal adenomas: An ex vivo and a clinical prospective feasibility study (SHARP trial).

机构信息

Division of Endoscopy, Shizuoka Cancer Center, Sunto-gun, Shizuoka, Japan.

Division of Pathology, Shizuoka Cancer Center, Sunto-gun, Shizuoka, Japan.

出版信息

J Gastroenterol Hepatol. 2024 Apr;39(4):667-673. doi: 10.1111/jgh.16452. Epub 2023 Dec 27.

Abstract

BACKGROUND AND AIM

Hot snare polypectomy using blend or coagulation current is widely used; however, it causes deeper tissue heat injury, leading to adverse events. We hypothesized that hot polypectomy using low-power pure cut current (PureCut, effect 1 10 W) could reduce deeper tissue heat injury. We conducted animal experiments to evaluate the deeper tissue heat injury and conducted a prospective clinical study to examine its cutting ability.

METHODS

In a porcine rectum, hot polypectomy using Blend current (EndoCut, effect 3 40 W) and low-power pure cut current was performed. The deepest part of heat destruction and thickness of the non-burned submucosal layer were evaluated histologically. Based on the results, we performed low-power pure cut current hot polypectomy for 10-14 mm adenoma. The primary endpoint was complete resection defined as one-piece resection with negative for adenoma in quadrant biopsies from the defect margin.

RESULTS

In experiments, all low-power pure-cut resections were limited within the submucosal layer whereas blend current resections coagulated the muscular layer in 13% (3/23). The remaining submucosal layer was thicker in low-power pure cut current than in blend current resections. In the clinical study, low-power pure-cut hot polypectomy removed all 100 enrolled polyps. For 98 pathologically neoplastic polyps, complete resection was achieved in 84 (85.7%, 95% confidence interval, 77-92%). The lower limit of the 95% confidence interval was not more than 15% below the pre-defined threshold of 86.6%. No severe adverse events occurred.

CONCLUSIONS

A novel low-power pure-cut hot polypectomy may be feasible for adenoma measuring 10-14 mm. (UMIN000037678).

摘要

背景与目的

使用混合或凝固电流的热圈套息肉切除术被广泛应用;然而,它会导致更深层的组织热损伤,从而引发不良事件。我们假设使用低功率纯切割电流(PureCut,效果 1 10 W)进行热息肉切除术可以减少更深层的组织热损伤。我们进行了动物实验来评估更深层的组织热损伤,并进行了一项前瞻性临床研究来检查其切割能力。

方法

在猪直肠中,使用 Blend 电流(EndoCut,效果 3 40 W)和低功率纯切割电流进行热圈套息肉切除术。通过组织学评估热破坏的最深部位和未烧伤的黏膜下层的厚度。基于这些结果,我们对 10-14 mm 的腺瘤进行了低功率纯切割电流热圈套切除术。主要终点是完全切除,定义为从缺损边缘的象限活检中,无腺瘤的整块切除。

结果

在实验中,所有低功率纯切割切除都局限于黏膜下层,而 Blend 电流切除则使 13%(3/23)的肌层凝固。低功率纯切割电流的残留黏膜下层比 Blend 电流切除更厚。在临床研究中,低功率纯切割热圈套切除术切除了所有 100 个入组的息肉。对于 98 个病理上为肿瘤性的息肉,84 个(85.7%,95%置信区间,77-92%)达到了完全切除。95%置信区间的下限不超过 86.6%预定阈值的 15%以下。没有发生严重的不良事件。

结论

一种新的低功率纯切割热圈套切除术对于 10-14 mm 的腺瘤可能是可行的。(UMIN000037678)。