Hunger Sandra, Rotsch Christian, Günther Florian, Drossel Welf-Guntram, Heyde Christoph-Eckhard
Fraunhofer Institute for Machine Tools and Forming Technology IWU, 01187 Dresden, Germany.
Department of Orthopaedic, Trauma and Plastic Surgery Clinic, University of Leipzig Medical Center, 04103 Leipzig, Germany.
Bioengineering (Basel). 2024 Feb 9;11(2):170. doi: 10.3390/bioengineering11020170.
In revision operations, ceramic heads of modular hip implants can be replaced. As the surface of the stem taper can be damaged, additional adapter sleeves are applied. The components are usually connected manually by the surgeon in a one-step procedure by hammer impacts. In this study, we investigated a two-step joining procedure with reproducible impaction force. First, the adapter sleeve and head were joined quasi-statically with a force of 2 kN using an assembly device. In the second step, these components were applied to the stem taper using a pulse-controlled instrument. For reference, the joints were assembled according to standard conditions using a tensile testing machine. An average pull-off force of 1309 ± 201 N was achieved for the components joined by the instrument, and the average measured values for the components joined by the testing machine were 1290 ± 140 N. All specimens achieved a force >350 N when released and therefore met the acceptance criterion defined for this study. This study showed that a modified procedure in two steps with a defined force has a positive effect on the reproducibility of the measured joining forces compared to previous studies.
在翻修手术中,可以更换模块化髋关节植入物的陶瓷头。由于柄部锥度表面可能受损,因此要使用额外的转接套筒。这些组件通常由外科医生通过锤击以一步法手动连接。在本研究中,我们研究了一种具有可重复冲击力的两步连接程序。首先,使用装配装置以2 kN的力将转接套筒和陶瓷头准静态连接。在第二步中,使用脉冲控制仪器将这些组件安装到柄部锥度上。作为参考,使用拉伸试验机按照标准条件组装接头。通过仪器连接的组件的平均拉脱力为1309±201 N,通过试验机连接的组件的平均测量值为1290±140 N。所有试样在释放时的力均> 350 N,因此符合本研究定义的验收标准。与先前的研究相比,本研究表明,采用两步法并施加确定的力的改良程序对测量的连接力的可重复性具有积极影响。