To evaluate the in vitro wettability and coefficient of friction of a novel amphiphilic polymeric surfactant (APS), poly(oxyethylene)-co-poly(oxybutylene) (PEO-PBO) releasing silicone hydrogel (SiHy) contact lens material (serafilcon A), compared to other reusable SiHy lens materials.
The release of fluorescently-labelled nitrobenzoxadiazole (NBD)-PEO-PBO was evaluated from serafilcon A over 7 days in a vial. The wettability and coefficient of friction of serafilcon A and three contemporary SiHy contact lens materials (senofilcon A; samfilcon A; comfilcon A) were evaluated using an in vitro blink model over their recommended wearing period; t = 0, 1, 7, 14 days for all lens types and t = 30 days for samfilcon A and comfilcon A (n = 4). Sessile drop contact angles were determined and in vitro non-invasive keratographic break-up time (NIKBUT) measurements were assessed on a blink model via the OCULUS Keratograph 5 M. The coefficient of friction was measured using a nano tribometer.
The relative fluorescence of NBD-PEO-PBO decreased in serafilcon A by approximately 18 % after 7 days. The amount of NBD-PEO-PBO released on day 7 was 50 % less than the amount released on day 1 (6.5±1.0 vs 3.4±0.5 µg/lens). The reduction in PEO-PBO in the lens also coincided with an increase in contact angles for serafilcon A after 7 days (p < 0.05), although there were no changes in NIKBUT or coefficient of friction (p > 0.05). The other contact lens materials had stable contact angles and NIKBUT over their recommended wearing period (p > 0.05), with the exception of samfilcon A, which had an increase in contact angle after 14 days as compared to t = 0 (p < 0.05). Senofilcon A and samfilcon A also showed an increase in coefficient of friction at 14 and 30 days, respectively, compared to their blister pack values (p < 0.05).
The results indicate that serafilcon A gradually depletes its reserve of PEO-PBO over 1 week, but this decrease did not significantly change the lens performance in vitro during this time frame.
评估新型两亲性高分子表面活性剂(APS)聚(氧乙烯)-共-聚(氧丁烯)(PEO-PBO)释放的硅水凝胶(SiHy)接触镜材料(帕利百)的体外润湿性和摩擦系数,与其他可重复使用的 SiHy 镜片材料相比。
在小瓶中评估荧光标记的硝苯并恶二唑(NBD)-PEO-PBO 从帕利百释放超过 7 天。使用体外眨眼模型评估帕利百和三种当代 SiHy 接触镜材料(新福康 A;山姆福康 A;康氟康 A)在推荐佩戴期内的润湿性和摩擦系数;对于所有镜片类型,t = 0、1、7、14 天,对于山姆福康 A 和康氟康 A,t = 30 天(n = 4)。通过 OCULUS Keratograph 5M 在眨眼模型上测定固着液滴接触角,并通过非侵入性角膜破裂时间(NIKBUT)测量评估体外无损伤角膜破裂时间(NIKBUT)。使用纳米摩擦计测量摩擦系数。
在 7 天后,帕利百中的 NBD-PEO-PBO 的相对荧光强度降低了约 18%。第 7 天释放的 NBD-PEO-PBO 量比第 1 天减少了 50%(6.5±1.0 比 3.4±0.5µg/镜片)。镜片中 PEO-PBO 的减少也与帕利百 7 天后接触角的增加相对应(p < 0.05),尽管 NIKBUT 或摩擦系数没有变化(p > 0.05)。在推荐的佩戴期内,其他接触镜材料的接触角和 NIKBUT 保持稳定(p > 0.05),除了山姆福康 A,与 t = 0 相比,14 天后接触角增加(p < 0.05)。与泡罩包装值相比,新福康 A 和山姆福康 A 还分别在 14 天和 30 天显示出摩擦系数的增加(p < 0.05)。