To determine the accuracy of the Reichert® Tono-Vera® Vet rebound tonometer for canine intraocular pressure (IOP) measurement.
Five normal canine ex vivo globes.
The anterior chambers of five freshly enucleated normal canine eyes were cannulated and connected to a reservoir of Plasma-Lyte A and a manometer. Starting at a manometric IOP of 5 mmHg, the pressure was progressively increased to 80 mmHg by raising the reservoir. Triplicate IOP measurements were taken with the Tono-Vera® Vet from the central cornea using the dog setting and compared to the manometric pressure by linear regression analysis and Bland-Altman plots.
There was a strong positive linear regression trend when comparing central corneal Tono-Vera® Vet IOPs to manometric pressures (r = .99) with solid agreement between the two methods. Compared to manometric IOPs, the Tono-Vera® Vet underestimated IOPs at higher pressures ≥70 mmHg.
Measurement of IOPs from the central cornea with the Tono-Vera® Vet provided accurate results over a large range in normal canine globes compared to direct manometry. The mild to moderate underestimation of IOPs at high pressures was not considered clinically relevant.
确定 Reichert® Tono-Vera® Vet 回弹眼压计测量犬眼内压(IOP)的准确性。
5 只正常犬离体眼球。
将 5 只新鲜眼球的前房进行套管穿刺,并与 Plasma-Lyte A 储液器和压力计连接。从压力计 IOP 为 5mmHg 开始,通过升高储液器将压力逐渐升高至 80mmHg。使用犬设置从中央角膜对 Tono-Vera® Vet 进行 3 次眼压测量,并通过线性回归分析和 Bland-Altman 图将其与压力计压力进行比较。
当将中央角膜 Tono-Vera® Vet IOP 与压力计压力进行比较时,存在强烈的正线性回归趋势(r=.99),两种方法之间具有良好的一致性。与压力计 IOP 相比,Tono-Vera® Vet 在较高压力(≥70mmHg)下低估了 IOP。
与直接压力计相比,Tono-Vera® Vet 从正常犬眼球中央角膜测量 IOP 可提供大范围的准确结果。在较高压力下对 IOP 的轻度至中度低估在临床上被认为无意义。