Suppr超能文献

验证贝克曼库尔特 AU5800 和 DxI800 自动化分析仪的钾、磷、镁、AST、LDH、铁、CA19-9 和维生素 D 的非报告溶血指数。

Verifying the nonreporting hemolysis index for potassium, phosphate, magnesium, AST, LDH, iron, CA 19-9, and vitamin D, using Beckman Coulter AU5800 and DxI800 automated analyzers.

机构信息

Department of Clinical Biochemistry, North Bristol NHS Trust-Southmead Hospital, Bristol, UK.

出版信息

Lab Med. 2024 Sep 4;55(5):624-626. doi: 10.1093/labmed/lmae027.

Abstract

BACKGROUND

Hemolysis is a common reason for nonreporting results in biochemistry and is measured using the hemolysis index (HI), with nonreporting limits set for analytes by manufacturers.

OBJECTIVE

To verify the nonreporting HI limit for potassium, phosphate, magnesium, aspartate aminotransferase (AST), lactate dehydrogenase (LDH), iron, CA19-9, and vitamin D on the Beckman Coulter AU5800/DxI800 analyzers.

METHOD

Hemolysate was created from EDTA-lined tubes of whole blood using an osmotic shock procedure. The hemolysate underwent serial dilutions with saline and was spiked in paired serum. The delta changes in HI and analyte concentration were measured, assessed using regression analysis, and compared against calculated reference change values.

RESULTS

A linear relationship between increasing HI and increasing analyte concentration (R2 > 0.9) was observed for potassium (y = 0.8864x), phosphate (y = 0.1079x), magnesium (y = 0.0678x), AST (y = 29.035x), and LDH (y = 350x). Increasing HI values did not have a linear effect on iron (y = -0.2544x), CA19-9 (y = 2.7019x), or vitamin D (y = 8.036x) concentrations.

CONCLUSION

The results from this experiment support increasing the HI nonreporting limit to 100 mg/dL for potassium; 200 mg/dL for magnesium; and 300 mg/dL for phosphate, CA19-9, and vitamin D. The iron assay is not affected by hemolysis as high as 500 mg/dL. The current HI nonreporting limit of 50 mg/dL is appropriate for LDH.

摘要

背景

溶血是导致生化检验结果无法报告的常见原因,通常使用溶血指数(HI)进行测量,制造商为各分析物设定了非报告限值。

目的

验证贝克曼库尔特 AU5800/DxI800 分析仪上钾、磷酸盐、镁、天门冬氨酸氨基转移酶(AST)、乳酸脱氢酶(LDH)、铁、CA19-9 和维生素 D 的非报告 HI 限值。

方法

使用渗透休克程序从 EDTA 线采血管的全血中制备溶血样本。溶血样本用生理盐水进行连续稀释,并与配对血清混合。测量 HI 和分析物浓度的差值变化,使用回归分析进行评估,并与计算的参考变化值进行比较。

结果

钾(y = 0.8864x)、磷酸盐(y = 0.1079x)、镁(y = 0.0678x)、AST(y = 29.035x)和 LDH(y = 350x)的 HI 和分析物浓度之间呈线性关系(R2 > 0.9)。HI 值增加对铁(y = -0.2544x)、CA19-9(y = 2.7019x)和维生素 D(y = 8.036x)浓度没有线性影响。

结论

本实验结果支持将 HI 的非报告限值提高到钾 100mg/dL;镁 200mg/dL;磷酸盐、CA19-9 和维生素 D 300mg/dL。铁检测不受高达 500mg/dL 的溶血影响。目前 50mg/dL 的 HI 非报告限值适用于 LDH。