Impact of wall thickness on the tissue cooling effect of cryoballoon ablation.

机构信息

Kansai Rosai Hospital Cardiovascular Center, 3-1-69 Inabaso, Amagasaki, Hyogo 660-8511, Japan.

出版信息

Europace. 2024 May 2;26(5). doi: 10.1093/europace/euae135.

Abstract

AIMS

Understanding of the tissue cooling properties of cryoballoon ablation during pulmonary vein (PV) isolation is lacking. The purpose of this study was to delineate the depth of the tissue cooling effect during cryoballoon freezing at the pulmonary venous ostium.

METHODS AND RESULTS

A left atrial-PV model was constructed using a three-dimensional printer with data from a patient to which porcine thigh muscle of various thicknesses could be affixed. The model was placed in a 37°C water tank with a PV water flow at a rate that mimicked biological blood flow. Cryofreezing at the PV ostium was performed five times each for sliced porcine thigh muscle of 2, 4, and 6 mm thickness, and sliced muscle cooling on the side opposite the balloon was monitored. The cooling effect was assessed using the average temperature of 12 evenly distributed thermocouples covering the roof region of the left superior PV. Tissue cooling effects were in the order of the 2, 4, and 6 mm thicknesses, with an average temperature of -41.4 ± 4.2°C for 2 mm, -33.0 ± 4.0°C for 4 mm, and 8.0 ± 8.7°C for 6 mm at 180 s (P for trend <0.0001). In addition, tissue temperature drops were steeper in thin muscle (maximum temperature drop per 5 s: 5.2 ± 0.9°C, 3.9 ± 0.7°C, and 1.3 ± 0.7°C, P for trend <0.0001).

CONCLUSION

The cooling effect of cryoballoon freezing is weaker in the deeper layers. Cryoballoon ablation should be performed with consideration to myocardial thickness.

摘要

目的

对于肺静脉(PV)隔离期间冷冻球囊消融的组织冷却特性缺乏了解。本研究旨在描绘冷冻球囊在肺静脉口冷冻时组织冷却效果的深度。

方法和结果

使用来自患者的三维打印机数据构建左心房-PV 模型,该模型可固定各种厚度的猪大腿肌肉。将模型放置在 37°C 的水箱中,其中 PV 水流速度模拟生物血流。对 2、4 和 6mm 厚的猪大腿肌肉切片各进行五次冷冻,监测球囊对面的切片肌肉冷却情况。使用覆盖左肺上腔静脉屋顶区域的 12 个均匀分布热电偶的平均温度评估冷却效果。组织冷却效果按 2、4 和 6mm 厚度的顺序排列,在 180 秒时 2mm 厚度的平均温度为-41.4±4.2°C,4mm 厚度的平均温度为-33.0±4.0°C,6mm 厚度的平均温度为 8.0±8.7°C(趋势 P<0.0001)。此外,在薄肌肉中组织温度下降更为陡峭(每 5 秒最大温度下降:5.2±0.9°C、3.9±0.7°C 和 1.3±0.7°C,趋势 P<0.0001)。

结论

冷冻球囊冷冻的冷却效果在较深的部位较弱。冷冻球囊消融时应考虑心肌厚度。