JMIR Mhealth Uhealth. 2024 Jun 3;12:e53964. doi: 10.2196/53964.
Due to aging of the population, the prevalence of aortic valve stenosis will increase drastically in upcoming years. Consequently, transcatheter aortic valve implantation (TAVI) procedures will also expand worldwide. Optimal selection of patients who benefit with improved symptoms and prognoses is key, since TAVI is not without its risks. Currently, we are not able to adequately predict functional outcomes after TAVI. Quality of life measurement tools and traditional functional assessment tests do not always agree and can depend on factors unrelated to heart disease. Activity tracking using wearable devices might provide a more comprehensive assessment.
This study aimed to identify objective parameters (eg, change in heart rate) associated with improvement after TAVI for severe aortic stenosis from a wearable device.
In total, 100 patients undergoing routine TAVI wore a Philips Health Watch device for 1 week before and after the procedure. Watch data were analyzed offline-before TAVI for 97 patients and after TAVI for 75 patients.
Parameters such as the total number of steps and activity time did not change, in contrast to improvements in the 6-minute walking test (6MWT) and physical limitation domain of the transformed WHOQOL-BREF questionnaire.
These findings, in an older TAVI population, show that watch-based parameters, such as the number of steps, do not change after TAVI, unlike traditional 6MWT and QoL assessments. Basic wearable device parameters might be less appropriate for measuring treatment effects from TAVI.
由于人口老龄化,主动脉瓣狭窄的患病率将在未来几年内大幅增加。因此,经导管主动脉瓣植入术(TAVI)的全球应用也将不断扩大。选择能改善症状和预后的患者至关重要,因为 TAVI 并非没有风险。目前,我们还不能充分预测 TAVI 后的功能结果。生活质量测量工具和传统的功能评估测试并不总是一致的,并且可能取决于与心脏病无关的因素。使用可穿戴设备进行活动跟踪可能提供更全面的评估。
本研究旨在从可穿戴设备中确定与严重主动脉瓣狭窄 TAVI 后改善相关的客观参数(例如心率变化)。
共 100 例接受常规 TAVI 的患者在手术前和手术后一周佩戴飞利浦 Health Watch 设备。在 TAVI 之前对 97 例患者和 TAVI 之后对 75 例患者进行了 Watch 数据分析。
与 6 分钟步行测试(6MWT)和 WHOQOL-BREF 问卷的身体限制域改善相反,总步数和活动时间等参数没有变化。
这些在较老的 TAVI 人群中的发现表明,与传统的 6MWT 和 QoL 评估不同,基于手表的参数(如步数)在 TAVI 后不会改变。基本的可穿戴设备参数可能不太适合衡量 TAVI 的治疗效果。