Sherbrooke Connectivity Imaging Laboratory (SCIL), Université de Sherbrooke, Sherbrooke, Canada.
MR Clinical Science, Philips Healthcare Canada, Mississauga, Canada.
Magn Reson Med. 2024 Nov;92(5):2207-2221. doi: 10.1002/mrm.30195. Epub 2024 Jun 24.
To fully characterize the orientation dependence of magnetization transfer (MT) and inhomogeneous MT (ihMT) measures in the whole white matter (WM), for both single-fiber and crossing-fiber voxels.
A characterization method was developed using the fiber orientation obtained from diffusion MRI (dMRI) with diffusion tensor imaging (DTI) and constrained spherical deconvolution. This allowed for characterization of the orientation dependence of measures in all of WM, regardless of the number of fiber orientation in a voxel. Furthermore, the orientation dependence inside 31 different WM bundles was characterized to evaluate the homogeneity of the effect. Variation of the results within and between-subject was assessed from a 12-subject dataset.
Previous results for single-fiber voxels were reproduced and a novel characterization was produced in voxels of crossing fibers, which seems to follow trends consistent with single-fiber results. Heterogeneity of the orientation dependence across bundles was observed, but homogeneity within similar bundles was also highlighted. Differences in behavior between MT and ihMT measures, as well as the ratio and saturation versions of these, were noted.
Orientation dependence characterization was proven possible over the entirety of WM. The vast range of effects and subtleties of the orientation dependence on MT measures showed the need for, but also the challenges of, a correction method.
全面描述单纤维和交叉纤维体素中整个白质(WM)中磁化转移(MT)和不均匀磁化转移(ihMT)测量的方向依赖性。
使用从扩散 MRI(dMRI)与扩散张量成像(DTI)和约束球分解获得的纤维方向开发了一种特征描述方法。这允许在不考虑体素中纤维方向数量的情况下,对所有 WM 中的测量的方向依赖性进行特征描述。此外,对 31 个不同 WM 束内的方向依赖性进行了特征描述,以评估效果的均匀性。从 12 个受试者数据集评估了受试者内和受试者间结果的变化。
重现了单纤维体素的先前结果,并对交叉纤维体素产生了新的特征描述,该描述似乎遵循与单纤维结果一致的趋势。观察到跨束的方向依赖性存在异质性,但也强调了类似束内的均匀性。注意到 MT 和 ihMT 测量之间的行为差异,以及这些测量的比值和饱和版本。
证明了在整个 WM 中进行方向依赖性特征描述是可行的。MT 测量的方向依赖性的广泛影响和细微差别表明需要但也存在校正方法的挑战。