Universidade Federal do Rio Grande do Sul - Grupo Biomec, Rua Felizardo, 750, bairro Jardim Botânico, Porto Alegre, Rio Grande do Sul, Brasil.
Eur Spine J. 2024 Oct;33(10):3798-3805. doi: 10.1007/s00586-024-08410-7. Epub 2024 Jul 17.
To assess the repeatability, intra and interrater reproducibility of the DIPA-S eHealth© system for capturing and measuring clinical variables of scoliosis, including frontal trunk imbalance (FTI), sagittal trunk imbalance (STI), and angle of trunk rotation (ATR).
Patients were photographed using the DIPA-S eHealth Capture© mobile application by family members, physiotherapists, or surgeons. Three photos were taken in each position: standing in the frontal and sagittal planes and in the axial plane in forward bending position of the trunk. The photos were analyzed by three independent evaluators using the DIPA-S eHealth Analysis© software. For repeatability, each photo was analyzed twice consecutively by the three evaluators. For intrarater reproducibility, only one evaluator reanalyzed the first photo from each plane with a 5-day interval. For interrater reproducibility, the three evaluators analyzed the first photo from each plane. The intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were used (p < 0.05).
The sample comprised 262 images (n = 30). Repeatability was excellent for all three evaluators in all three planes (ICC 0.94 to 1.00). Intrarater reproducibility was excellent in all three planes (ICC 0.88 to 0.99). Interrater analysis was excellent for the frontal and axial planes (ICC 0.98 and 0.93), respectively. However, it was weak in the sagittal plane ICC 0.32. The SEM ranged from 0.02 to 17.6 and MDC from 0.03 to 34.6.
The DIPA-S eHealth© Capture and Analysis system demonstrates reproducibility for use in the clinical assessment of scoliosis through teleconsultations.
评估 DIPA-S 电子健康系统捕捉和测量脊柱侧凸临床变量(包括额状面躯干不平衡(FTI)、矢状面躯干不平衡(STI)和躯干旋转角度(ATR))的重复性、内部和外部评估者间的可重复性。
由家庭成员、物理治疗师或外科医生使用 DIPA-S 电子健康 Capture©移动应用程序为患者拍照。在每个位置拍摄三张照片:站立在额状面和矢状面以及躯干前屈位的轴向平面。使用 DIPA-S 电子健康 Analysis©软件,由三位独立评估者分析照片。对于重复性,每位评估者连续两次分析三张照片。对于内部评估者间的可重复性,只有一位评估者在 5 天间隔后重新分析每个平面的第一张照片。对于外部评估者间的可重复性,三位评估者分析每个平面的第一张照片。使用组内相关系数(ICC)、测量标准误差(SEM)和最小可检测变化(MDC)(p<0.05)。
该样本包括 262 张图像(n=30)。在所有三个平面上,所有三位评估者的重复性均为优秀(ICC 0.94 至 1.00)。在所有三个平面上,内部评估者间的可重复性均为优秀(ICC 0.88 至 0.99)。对于额状面和轴向平面,外部评估者间的分析分别为优秀(ICC 0.98 和 0.93)。然而,在矢状面 ICC 为 0.32,评估结果为弱。SEM 范围为 0.02 至 17.6,MDC 范围为 0.03 至 34.6。
DIPA-S 电子健康 Capture 和 Analysis 系统在通过远程咨询对脊柱侧凸的临床评估中具有可重复性。