Department of Radiation, The Affiliated Cancer Hospital of Zhengzhou University & Henan Cancer Hospital, Zhengzhou, China.
Technol Cancer Res Treat. 2024 Jan-Dec;23:15330338241264848. doi: 10.1177/15330338241264848.
To investigate the effect of various frequencies of bolus use on the superficial dose of volumetric modulated arc therapy after modified radical mastectomy for breast cancer.
Based on the computed tomography images of a female anthropomorphic breast phantom, a 0.5 cm silicone-based 3D-printed bolus was created. Nine points evenly distributed on the breast skin were selected for assessing the skin dose, and a volume of subcutaneous lymphatic drainage of the breast (noted as ROI2-3) was delineated for assessing the chest wall dose. The treatment plans with and without bolus (plan_wb and plan_nb) were separately designed using the prescription of 50 Gy in 25 fractions following the standard dose constraints of the adjacent organ at risk. To characterize the accuracy of treatment planning system (TPS) dose calculations, the doses of the nine points were measured five times by thermoluminescence dosimeters (TLDs) and then were compared with the TPS calculated dose.
Compared with Plan_nb (144.46 ± 10.32 cGy), the breast skin dose for plan_wb (208.75 ± 4.55 cGy) was significantly increased (t = -18.56, < 0.001). The deviation of skin dose was smaller for Plan_wb, and the uniformity was significantly improved. The calculated value of TPS was in good agreement with the measured value of TLD, and the maximum deviation was within 5%. Skin and ROI2-3 doses were significantly increased with increasing frequencies of bolus applications. The mean dose of the breast skin and ROI2-3 for 15 and 23 times bolus applications were 45.33 Gy, 50.88 Gy and 50.36 Gy, 52.39 Gy, respectively.
3D printing bolus can improve the radiation dose and the accuracy of the planned dose. Setting Plan_wb to 15 times for T1-3N+ breast cancer patients and 23 times for T4N+ breast cancer patients can meet the clinical need. Quantitative analysis of the bolus application frequency for different tumor stages can provide a reference for clinical practice.
研究不同剂量率的推注对乳腺癌改良根治术后容积调强弧形治疗(VMAT)体表剂量的影响。
基于女性人体乳房体模的计算机断层扫描图像,制作了 0.5 cm 厚的硅基 3D 打印体模。选择 9 个均匀分布在乳房皮肤上的点来评估皮肤剂量,勾画乳房皮下淋巴引流区(ROI2-3)以评估胸壁剂量。根据相邻危及器官的标准剂量限制,分别为处方 50 Gy/25 次设计有体模(plan_wb)和无体模(plan_nb)的治疗计划。为了评估治疗计划系统(TPS)剂量计算的准确性,使用热释光剂量计(TLD)五次测量这 9 个点的剂量,然后与 TPS 计算的剂量进行比较。
与 plan_nb(144.46±10.32 cGy)相比,plan_wb(208.75±4.55 cGy)的乳房皮肤剂量显著增加(t=-18.56,<0.001)。有体模的计划的皮肤剂量偏差较小,均匀性显著提高。TPS 的计算值与 TLD 的实测值吻合良好,最大偏差在 5%以内。随着推注次数的增加,皮肤和 ROI2-3 剂量显著增加。15 次和 23 次推注应用时乳房皮肤和 ROI2-3 的平均剂量分别为 45.33 Gy、50.88 Gy 和 50.36 Gy、52.39 Gy。
3D 打印体模可以提高照射剂量和计划剂量的准确性。对于 T1-3N+乳腺癌患者,将 plan_wb 设置为 15 次,对于 T4N+乳腺癌患者,将 plan_wb 设置为 23 次,可以满足临床需要。对不同肿瘤分期的推注应用频率进行定量分析,可以为临床实践提供参考。