Department of Cardiology, Fu Jen Catholic University Hospital.
R&D Software Department, AViTA Corporation, New Taipei City, Taiwan.
Blood Press Monit. 2024 Dec 1;29(6):321-326. doi: 10.1097/MBP.0000000000000722. Epub 2024 Aug 9.
To determine the accuracy of the AViTA BP636 upper-arm home blood pressure (BP) monitor in general populations with small to extra-large arms according to the ANSI/AAMI/ISO Universal Standard (ISO 81060-2 : 2018/Amd1 : 2020).
The AViTA BPM636 was utilized on 85 adult participants with upper-arm circumferences ranging from 15 to 52 cm, and four cuffs of different sizes (small: 15-22 cm, normal: 22-33 cm, large: 33-42 cm, and extra large: 42-52 cm) were used in this study. The recruitment fulfills the criteria of the ANSI/AAMI/ISO Universal Standard 81060-2 : 2018/Amd1 : 2020 in the general population using the same arm sequential BP measurement method.
For validation criterion 1, the mean ± SD of the differences between the test device and reference blood pressure readings was -0.35 ± 4.68/0.56 ± 3.93 mmHg (systolic/diastolic) for adults. The mean difference of SBP between the test device and the reference device was -0.35 mmHg with a SD of 3.50 mmHg, which was within the acceptable range for criterion 2 (i.e. less than 6.93 mmHg) Similarly, the mean difference in DBP was 0.56 mmHg with a SD of 3.38 mmHg, which also met the criterion 2 (i.e. within 6.91 mmHg).
AViTA BPM636 passed the criteria of the ANSI/AAMI/ISO 81060-2 : 2018/Amd 1 : 2020 protocol. It can be recommended for home-based BP measurements in the general population and people with extra-large arms ranging from 42 to 52 cm.
根据美国国家标准协会/美国医疗器械促进协会/国际标准化组织通用标准(ISO 81060-2:2018/Amd1:2020),确定 AViTA BP636 上臂家庭血压(BP)监测仪在小到特大臂围的一般人群中的准确性。
在 85 名成年参与者中使用 AViTA BPM636,上臂周长范围为 15 至 52cm,使用了四种不同尺寸的袖带(小号:15-22cm,中号:22-33cm,大号:33-42cm,特大号:42-52cm)。本研究符合美国国家标准协会/美国医疗器械促进协会/国际标准化组织通用标准 81060-2:2018/Amd1:2020 在一般人群中使用相同手臂顺序血压测量方法的标准。
对于验证标准 1,测试设备和参考血压读数之间差异的平均值±SD 为-0.35±4.68/0.56±3.93mmHg(收缩压/舒张压)成人。测试设备和参考设备之间 SBP 的平均差异为-0.35mmHg,标准差为 3.50mmHg,符合标准 2(即小于 6.93mmHg)的要求。同样,DBP 的平均差异为 0.56mmHg,标准差为 3.38mmHg,也符合标准 2(即小于 6.91mmHg)的要求。
AViTA BPM636 通过了美国国家标准协会/美国医疗器械促进协会/国际标准化组织 81060-2:2018/Amd 1:2020 协议的标准。它可以推荐用于一般人群和臂围在 42 至 52cm 之间的特大人群的家庭血压测量。