Rizzieri Nicola, Facchin Alessio
Optometry and Vision Science, University of Latvia, LV-1586 Riga, Latvia.
Rizzieri Vision Center, 46041 Asola, Italy.
Vision (Basel). 2024 Jul 26;8(3):46. doi: 10.3390/vision8030046.
This study evaluates the inter-device measurement properties of partial coherence interferometry (PCI) and spectral domain optical coherence tomography (SD-OCT) in measuring axial length, particularly for myopia management. We recruited 82 eyes from 41 adult participants with a mean age of 31.0 ± 17.6 years and a mean spherical equivalent of -2.20 ± 2.28 D. Axial length was measured using SD-OCT and PCI for both the right and left eyes. Agreement between the two measurements was assessed using Bland-Altman analysis, and graphs and values were compared with linear mixed models. The results show a near-to-zero and non-significant bias between measurements. The 95% limits of agreement showed a value of 0.06 mm. Both devices can accurately measure the axial length. OCT biometry performed with SD-OCT can be successfully interchanged with partial coherence interferometry, but they should be cautiously interchanged when performing longitudinal comparisons.
本研究评估了部分相干干涉测量法(PCI)和频域光学相干断层扫描(SD-OCT)在测量眼轴长度方面的设备间测量特性,特别是用于近视管理。我们从41名成年参与者中招募了82只眼睛,这些参与者的平均年龄为31.0±17.6岁,平均等效球镜度数为-2.20±2.28 D。使用SD-OCT和PCI对右眼和左眼进行眼轴长度测量。使用Bland-Altman分析评估两次测量之间的一致性,并将图表和数值与线性混合模型进行比较。结果显示测量之间的偏差接近零且无统计学意义。95%一致性界限显示值为0.06 mm。两种设备都能准确测量眼轴长度。使用SD-OCT进行的OCT生物测量可以成功地与部分相干干涉测量法互换,但在进行纵向比较时应谨慎互换。