Terzioğlu Hakan, Sarıkaya Can Hakan, Gürbüz Rıza, Karaağaçlıoğlu Lale
Department of Prosthodontics, Faculty of Dentistry, Ankara University, Ankara, Turkey.
Department of Prosthodontics, Graduate School of Health Sciences, Ankara University, Ankara, Turkey.
Case Rep Dent. 2025 Mar 18;2025:4064452. doi: 10.1155/crid/4064452. eCollection 2025.
The aim of this study was to determine the causes of fatigue and implant failures in an implant-retained overdenture case. One out of four implants portrayed inadequate osseointegration during healing abutment placement, and one was fractured from the middle to the apical third after 1 year of denture use. Various aspects, such as insufficient cooling while implant socket preparation, incorrect prosthodontic planning, and erroneous design of the denture, were evaluated. Macroscopic and microscopic inspections of the fractured implant body were assessed by scanning electron microscopy (SEM) and optical microscopy. Consequently, metallic fatigue along with secondary cracks and beach marks were evident. It was deduced that the failed and fractured implants depicted substandard osseointegration; thus, they were susceptible to the wear out effect of the occlusal forces. Unfavorable osseointegration caused the fulcrum axis to move apically instead of forming on the neck part of the implant, and micromovements around the fulcrum axis mustered the fracture.
本研究的目的是确定种植体支持的覆盖义齿病例中疲劳和种植体失败的原因。在四颗种植体中,有一颗在愈合基台放置期间表现出骨结合不足,另一颗在义齿使用1年后从中间至根尖三分之一处发生骨折。对种植窝制备时冷却不足、义齿修复计划不正确以及义齿设计错误等各个方面进行了评估。通过扫描电子显微镜(SEM)和光学显微镜对骨折种植体主体进行了宏观和微观检查。结果,明显可见金属疲劳以及二次裂纹和海滩痕迹。据推断,失败和骨折的种植体表现出骨结合不达标;因此,它们易受咬合力磨损效应的影响。不利的骨结合导致支点轴向根尖移动,而不是在种植体颈部形成,围绕支点轴的微动促使了骨折的发生。