在超高分辨率(7特斯拉)磁共振成像环境中对镍钛诺乳腺活检夹进行安全许可和伪影测试

Safety Clearance and Artifact Testing of a Nitinol Breast Biopsy Clip in an Ultra-High Resolution (7 Tesla) Magnetic Resonance Imaging Environment.

作者信息

Dong William, Ramchandran Kanchna, Galloy Adam E, Nino Marco A, Kleingartner Marla, Piaszynski John, Raghavan Madhavan L, Phadke Sneha, Magnotta Vincent A

机构信息

Department of Internal Medicine, Carver College of Medicine, University of Iowa, Iowa City, Iowa, United States.

Roy J. Carver Department of Biomedical Engineering, College of Engineering, University of Iowa, Iowa City, Iowa, United States.

出版信息

Indian J Radiol Imaging. 2024 Oct 7;35(2):263-271. doi: 10.1055/s-0044-1791491. eCollection 2025 Apr.

Abstract

The lack of safety clearance of several metallic breast biopsy clips in 7 Tesla (T) poses a significant hurdle to using advanced magnetic resonance imaging (MRI) techniques in clinical management or cancer research.  This article assesses the Ultracor Twirl clip for safety and imaging artifacts in a 7T MRI scanner.  This study can be categorized as a phantom study.  Tests for magnetic susceptibility (translational attraction and torque), MRI-related heating, and artifacts were conducted based on the American Society for Testing and Materials standards. The magnetic susceptibility tests evaluated the scanner's magnetic force that can cause clip movement and rotation. The heating test was conducted with customized MRI parameters of short TR and maximum echo-train length, designed to induce temperature change. The artifact test, using T1-weighted spin and gradient echo imaging sequences, evaluated potential image misrepresentations (localized signal loss) caused by the clip's metallic properties.  None.  The magnetic susceptibility tests indicated no noticeable translational or rotational force exerted by the MRI scanner. The heating test indicated no significant temperature change (<0.3°C) in the testing gel when the clip was absent/present, both within the safety threshold (<1°C). The artifact test's clip images all contained an artifact (largest radius = 10.7 mm). These cumulative results indicate that this clip is safe in 7T scanners. Scanning at least 10.7 mm away from the clip avoids potential signal loss in the region of interest.

摘要

几种金属乳腺活检夹在7特斯拉(T)磁场下缺乏安全许可,这对在临床管理或癌症研究中使用先进的磁共振成像(MRI)技术构成了重大障碍。 本文评估了Ultracor Twirl夹在7T MRI扫描仪中的安全性和成像伪影。 本研究可归类为体模研究。 根据美国材料与试验协会标准,进行了磁敏感性(平移吸引力和扭矩)、MRI相关加热及伪影测试。磁敏感性测试评估了可能导致夹子移动和旋转的扫描仪磁力。加热测试采用定制的短TR和最大回波链长度的MRI参数进行,旨在诱导温度变化。伪影测试使用T1加权自旋和梯度回波成像序列,评估由夹子金属特性引起的潜在图像失真(局部信号丢失)。 无。 磁敏感性测试表明MRI扫描仪未施加明显的平移或旋转力。加热测试表明,夹子存在/不存在时,测试凝胶中的温度均无显著变化(<0.3°C),均在安全阈值(<1°C)范围内。伪影测试的夹子图像均包含伪影(最大半径 = 10.7 mm)。这些累积结果表明,该夹子在7T扫描仪中是安全的。在距离夹子至少10.7 mm处进行扫描可避免感兴趣区域出现潜在的信号丢失。