Richardson G H, Mortensen M S, Crockett R G
J Assoc Off Anal Chem. 1978 Sep;61(5):1038-40.
A vapor pressure osmometer (VPO) was successfully studied in 1975. An improved osmometer with greater range and temperature stability was evaluated. Repeatability on 290 mOs/kg standard solution was 289.5 +/- 1.08. The mean and standard deviation on 36 water-free milk samples was 280.5 +/- 4 compared to 280.1 +/- 3 in the previous study. Instrument data from 2 osmometers and 2 cryoscopes produced acceptable chi-square values, and the correlation coefficients between methods were better than 0.99 for samples containing 1--21% added water. In a survey involving 760 samples, more than 37% of producer samples in one cheese plant contained greater than 3% added water. Additional evidence of added water in market milk samples suggests the immediate application of the VPO to help prevent milk adulteration. The vapor pressure osomometer method for quantitating added water in milk has been adopted as official first action.
1975年对蒸汽压渗透压计(VPO)进行了成功研究。对一种量程更大且温度稳定性更高的改进型渗透压计进行了评估。在290毫渗摩尔/千克标准溶液上的重复性为289.5±1.08。36个无水牛奶样品的平均值和标准差为280.5±4,而之前的研究中为280.1±3。来自2台渗透压计和2台冰点渗透压计的仪器数据产生了可接受的卡方值,对于添加了1%至21%水分的样品,各方法之间的相关系数优于0.99。在一项涉及760个样品的调查中,一家奶酪厂超过37%的生产样品添加水分超过3%。市售牛奶样品中添加水分的更多证据表明应立即应用VPO来帮助防止牛奶掺假。用于定量牛奶中添加水分的蒸汽压渗透压计法已被采用为官方首次行动方法。