Flow-monitoring system for high-frequency ventilation.

作者信息

Hancock D E, Pinchak A C, Hagen J F

出版信息

Crit Care Med. 1984 Sep;12(9):824-6. doi: 10.1097/00003246-198409000-00033.

Abstract

With the addition of appropriate logic circuitry and a single-board microcomputer, a thermistor flowmeter becomes a clinically useful ventilation monitor. The microcomputer calculates upper and lower reference (voltage) levels from the baseline (normal) flow signal generated by the thermistor bridge. If the flow signal fails to pass repetitively through one or both of these reference levels, visual and audio alarms are generated. Tests performed with both high-frequency (3 Hz) and low-frequency (0.2 Hz) ventilator systems show the usefulness of this system in sensing a variety of malfunctions, including endotracheal tube occlusions, leaks, changes in flow or tidal volume, and hose disconnects.

摘要

通过添加适当的逻辑电路和单板微型计算机,热敏电阻流量计就成为临床上有用的通气监测器。微型计算机根据热敏电阻桥产生的基线(正常)流量信号计算出上下参考(电压)电平。如果流量信号不能反复通过其中一个或两个参考电平,就会发出视觉和听觉警报。在高频(3Hz)和低频(0.2Hz)通气系统上进行的测试表明,该系统在检测各种故障方面很有用,包括气管内导管堵塞、漏气、流量或潮气量变化以及软管断开。