使用MA-1呼吸机进行高频正压通气。

High-frequency positive-pressure ventilation with the MA-1 ventilator.

作者信息

Pinchak A C, Hagen J F, Hancock D, Kovijanic C

出版信息

Crit Care Med. 1984 Sep;12(9):810-4. doi: 10.1097/00003246-198409000-00030.

Abstract

A conventional, low-frequency ventilator was modified to ventilate dogs at high frequencies of 85 to 185 cycle/min while cardiovascular and pulmonary variables were monitored. Although gas transport was adequate, cardiac output was diminished when compared to low-frequency ventilation. The addition of an in-line pneumotachograph markedly increased PaCO2 during high-frequency ventilation. Carbon dioxide transport was primarily dependent upon the magnitude of the tidal volume at all high frequencies tested.

摘要

一台传统的低频呼吸机被改装,以便在85至185次/分钟的高频下为狗进行通气,同时监测心血管和肺部变量。尽管气体传输充足,但与低频通气相比,心输出量有所减少。在高频通气期间,增加一台在线式呼吸流速计会显著提高动脉血二氧化碳分压(PaCO2)。在所有测试的高频下,二氧化碳传输主要取决于潮气量的大小。