A critical improvement of the local thermodilution method for measuring renal blood flow in man.

作者信息

Magrini F, Guo-Quing L

出版信息

Cardiovasc Res. 1982 Jun;16(6):350-4. doi: 10.1093/cvr/16.6.350.

Abstract

A new catheter with a "pig tail" shaped end has been used for renal venous blood outflow measurement by the local thermodilution method. Left and right renal blood flow were measured in nine hypertensive patients with monolateral renal artery stenosis (previously documented by selective renal angiography). Percutaneous introduction and positioning of the "pig tail" catheter in the renal veins was performed without complications. In all cases no deformities in the dilution curves due to: a) contact of the thermistor with the vessel walls; b) shifting baseline during injection; or c) variable and in homogeneous mixing, were observed. Mean percentage variation of five consecutive renal blood flow determinations was 6% in a flow range 154 to 718 cm3 . min-1. We conclude that the use of "pig tail" catheters represents a critical improvement of the local thermodilution method for measuring renal blood flow.

摘要

一种带有“猪尾”形末端的新型导管已被用于通过局部热稀释法测量肾静脉血流。对9例单侧肾动脉狭窄的高血压患者(先前经选择性肾血管造影证实)测量了左右肾血流量。经皮将“猪尾”导管插入并定位在肾静脉中,未出现并发症。在所有病例中,均未观察到由于以下原因导致的稀释曲线变形:a)热敏电阻与血管壁接触;b)注射期间基线偏移;或c)可变且不均匀混合。在154至718 cm3·min-1的血流范围内,连续5次肾血流量测定的平均百分比变化为6%。我们得出结论,“猪尾”导管的使用代表了测量肾血流量的局部热稀释法的一项关键改进。