Suppr超能文献

使用表面肝素化导管电极进行连续动脉内氧分压监测。对传统血气分析的一致性以及非肝素化犬长期电极功能的研究。

Continuous intra-arterial PO2 monitoring with a surface heparinized catheter electrode. A study of conformity in conventional blood gas analysis and of long-term electrode function in the non-heparinized dog.

作者信息

Nilsson E, Edwall G, Larsson R, Olsson P

出版信息

Scand J Clin Lab Invest. 1982 Jun;42(4):331-8.

PMID:7134818
Abstract

Surface-heparinized catheter PO2 electrodes were used for continuous intra-arterial monitoring in non-heparinized dogs. During one-day experiments the read-out of the electrode monitoring unit was compared with that of conventional blood gas analysis. Furthermore, the electrodes were studied after long-term implantation. The intra-arterial PO2 electrodes have previously been reported to be linear in the gas phase, water phase and in heparinized blood. The present study demonstrates that the catheter electrodes remain linear after surface heparinization. As compared with the conventional blood gas analysis it was found that the desk analyser in the range PaO2 greater than 18 kPa underestimated the actual PaO2 as compared to the invasively monitored PaO2, but in the range below 7 kPa the desk analyser overestimated the PaO2. Surface-heparinized PO2 catheter electrodes were studied after up to 23 days of implantation. No coagulation phenomenon was found around the catheters, neither macroscopically nor by scanning electron microscopy. The electrical function withstood 7 days of implantation. After more extended periods the electrical leads were broken without any damage to the catheter itself. When compared with acutely implanted surface-heparinized electrodes, there was no difference in monitoring characteristics between the acutely implanted and the long-term implanted PO2 electrodes.

摘要

表面肝素化导管PO2电极用于未肝素化犬的连续动脉内监测。在为期一天的实验中,将电极监测单元的读数与传统血气分析的读数进行比较。此外,还对长期植入后的电极进行了研究。此前有报道称,动脉内PO2电极在气相、水相和肝素化血液中呈线性。本研究表明,导管电极表面肝素化后仍保持线性。与传统血气分析相比,发现台式分析仪在PaO2大于18 kPa的范围内,与有创监测的PaO2相比低估了实际PaO2,但在低于7 kPa的范围内,台式分析仪高估了PaO2。对表面肝素化PO2导管电极进行了长达23天的植入研究。在导管周围未发现凝血现象,无论是宏观上还是通过扫描电子显微镜观察均未发现。其电功能在植入7天后仍能维持。在更长时间后,电线断裂,但导管本身未受损。与急性植入的表面肝素化电极相比,急性植入和长期植入的PO2电极在监测特性上没有差异。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验