Suppr超能文献

一种用于控制麻醉机中氧气混合气体的简化概念——更好、更便宜且更便于用户使用?

A simplified concept for controlling oxygen mixtures in the anaesthetic machine--better, cheaper and more user-friendly?

作者信息

Berge J A, Gramstad L, Grimnes S

机构信息

Department of Clinical Engineering, National Hospital, University of Oslo, Norway.

出版信息

Acta Anaesthesiol Scand. 1995 May;39(4):563-7. doi: 10.1111/j.1399-6576.1995.tb04121.x.

Abstract

Modern anaesthetic machines are equipped with several safety components to prevent delivery of hypoxic mixtures. However, such a technical development has increased the complexity of the equipment. We report a reconstructed anaesthetic machine in which a paramagnetic oxygen analyzer has provided the means to simplify the apparatus. The new machine is devoid of several components conventionally included to prevent hypoxic mixtures: oxygen failure protection device, reservoir O2 alarm, N2O/air selector, and proportioning system for oxygen/nitrous oxide delivery. These devices have been replaced by a simple safety system using a paramagnetic oxygen analyzer at the common gas outlet, which in a feed-back system cuts off the supply of nitrous oxide whenever the oxygen concentration falls below 25%. The simplified construction of the anaesthetic machine has important consequences for safety, cost and user-friendliness. Reducing the complexity of the construction also simplifies the pre-use checkout procedure, and an efficient 5-point check list is presented for the new machine.

摘要

现代麻醉机配备了多个安全组件,以防止输送缺氧混合气。然而,这种技术发展增加了设备的复杂性。我们报告了一台经过改造的麻醉机,其中顺磁氧分析仪提供了简化该设备的方法。这台新机器没有几个传统上用于防止缺氧混合气的组件:氧气故障保护装置、储氧器氧气报警器、一氧化二氮/空气选择器以及氧气/一氧化二氮输送配比系统。这些装置已被一个简单的安全系统所取代,该系统在共用气体出口处使用顺磁氧分析仪,在反馈系统中,每当氧气浓度降至25%以下时,就切断一氧化二氮的供应。麻醉机结构的简化对安全性、成本和用户友好性都有重要影响。降低结构的复杂性也简化了使用前的检查程序,并且为这台新机器提供了一份有效的五点检查表。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验