Terada Y, Wanibuchi Y
Department of Cardiovascular Surgery, Mitsui Memorial Hospital, Tokyo, Japan.
Surg Today. 1994;24(9):823-5. doi: 10.1007/BF01636314.
A 58-year-old man required an emergency redo mitral valve replacement due to a rapidly deteriorating failure of the Carpentier-Edwards porcine bioprosthesis within [corrected] a few days after the initial operation. The removed bioprosthesis showed that its three commissures had all peeled off each stent, which may have occurred within a short time, and caused massive, rapidly progressive trans-bioprosthetic regurgitation.