Ducros V, Ruffieux D, Belin N, Favier A
Laboratoire de Biochimie C, Hôpital Michallon, Grenoble, France.
Analyst. 1994 Aug;119(8):1715-7. doi: 10.1039/an9941901715.
Two digestion methods for the determination of selenium in biological samples have been compared: a classic digestion in a heating block by means of wet acid-ashing (HNO3-HClO4) and a microwave digestion in an open-vessel system by heating under reflux with HNO3-H2O2. The two methods were used to determine Se in plasma samples and different reference materials. There was no difference between the mean results, and the correlation coefficient for wet acid-ashing versus microwave digestion was 0.94. The two methods yielded accurate results for the certified materials analysed.
一种是在加热块中通过湿酸灰化(硝酸 - 高氯酸)进行的经典消化,另一种是在开放容器系统中通过硝酸 - 过氧化氢回流加热进行的微波消化。这两种方法用于测定血浆样品和不同参考物质中的硒。平均结果之间没有差异,湿酸灰化与微波消化的相关系数为0.94。对于所分析的标准物质,这两种方法都得出了准确的结果。