Influence of infusion pump operation and flow rate on hemodynamic stability during epinephrine infusion.

作者信息

Klem S A, Farrington J M, Leff R D

机构信息

Department of Pediatrics, University of Kansas Medical Center, Kansas City.

出版信息

Crit Care Med. 1993 Aug;21(8):1213-7. doi: 10.1097/00003246-199308000-00023.

Abstract

OBJECTIVE

To determine whether variations in the flow rate of epinephrine solutions administered via commonly available infusion pumps lead to significant variations in blood pressure (BP) in vivo.

DESIGN

Prospective, randomized, crossover study with factorial design, using infusion pumps with four different operating mechanisms (pulsatile diaphragm, linear piston/syringe, cyclic piston-valve, and linear peristaltic) and three drug delivery rates (1, 5, and 10 mL/hr).

SUBJECTS

Two healthy, mixed-breed dogs (12 to 16 kg).

INTERVENTIONS

Dogs were made hypotensive with methohexital bolus and continuous infusion. BP was restored to normal with constant-dose epinephrine infusion via two pumps at each rate.

MEASUREMENTS

Femoral mean arterial pressure (MAP) was recorded every 10 secs. Pump-flow continuity was quantitated in vitro using a digital gravimetric technique. Variations in MAP and flow continuity were expressed by the coefficient of variation; analysis of variance was used for comparisons.

RESULTS

The mean coefficients of variations for MAP varied from 3.8 +/- 3.1% (linear piston/syringe) to 6.1 +/- 6.6% (linear peristaltic), and from 3.4 +/- 2.2% (10 mL/hr) to 7.9 +/- 6.6% (1 mL/hr). The coefficients of variation for in vitro flow continuity ranged from 9 +/- 8% (linear piston-syringe) to 250 +/- 162% (pulsatile diaphragm), and from 35 +/- 44% (10 mL/hr) to 138 +/- 196% (1 mL/hr). Both the type of pump and infusion rate significantly (p < .001) influenced variation in drug delivery rate. The 1 mL/hr infusion rate significantly (p < .01) influenced MAP variation. Cyclic fluctuations in MAP of < or = 30 mm Hg were observed using the pulsatile diaphragm pump at 1 mL/hr.

CONCLUSION

Factors inherent in the operating mechanisms of infusion pumps may result in clinically important hemodynamic fluctuations when administering a concentrated short-acting vasoactive medication at slow infusion rates.