We studied the response time (RT) and reliability of three neonatal patient-triggered ventilator (PTV) systems: the Draeger Babylog 8000, the Bear Cub enhancement module (CEM), and the Infrasonics Star Sync. In 10 adult rabbits, airway flow and pressure recordings showed the RT of the Star Sync to be shorter than that of the Bear CEM (53 +/- 13 versus 65 +/- 15 ms, p < 0.05), and both were shorter than that of the Babylog (95 +/- 24 ms, p < 0.01) by ANOVA. The RT of the Bear CEM and the Babylog increased significantly at decreased trigger sensitivity settings. All ventilators triggered successfully on assist-control (A/C). However, the Babylog had a higher rate of asynchrony on SIMV (30 +/- 25%) than the Bear CEM (1.1 +/- 0.3%) and the Star Sync (1.2 +/- 0.4%), p < 0.01. In 10 infants with respiratory failure, recordings of airway flow and pressure were made at ventilator inspiratory time (Ti) settings of 0.3, 0.4, and 0.5 s on assist-control and on SIMV at rates of 15, 30, 45, and 60 breaths/min. The Star Sync and Bear CEM triggered successfully on A/C (100%) and had low rates of asynchrony on SIMV (1 to 3%). The Babylog had a lower success rate on A/C (70 +/- 12%) and a higher rate of asynchrony on SIMV (29 +/- 30%) than the other two ventilators; p < 0.01. The lower reliability of the Babylog was due to its variable refractory period (0.2 to 0.5 s, to equal the set Ti).(ABSTRACT TRUNCATED AT 250 WORDS)
我们研究了三种新生儿患者触发通气系统(PTV)的反应时间(RT)和可靠性:德尔格Babylog 8000、熊牌增强模块(CEM)和英弗森星同步系统。在10只成年兔中,气道流量和压力记录显示,星同步系统的反应时间短于熊牌CEM(53±13毫秒对65±15毫秒,p<0.05),通过方差分析,两者均短于Babylog(95±24毫秒,p<0.01)。在触发灵敏度设置降低时,熊牌CEM和Babylog的反应时间显著增加。所有呼吸机在辅助控制(A/C)模式下均能成功触发。然而,Babylog在同步间歇指令通气(SIMV)模式下的不同步率(30±25%)高于熊牌CEM(1.1±0.3%)和星同步系统(1.2±0.4%),p<0.01。在10例呼吸衰竭婴儿中,在辅助控制和SIMV模式下,分别在15、30、45和60次/分钟的呼吸频率下,于呼吸机吸气时间(Ti)设置为0.3、0.4和0.5秒时记录气道流量和压力。星同步系统和熊牌CEM在A/C模式下成功触发(100%),在SIMV模式下不同步率较低(1%至3%)。Babylog在A/C模式下的成功率较低(70±12%),在SIMV模式下的不同步率高于其他两种呼吸机(29±30%);p<0.01。Babylog可靠性较低的原因是其可变不应期(0.2至0.5秒,以匹配设定的Ti)。(摘要截断于250字)