使用123I-碘美西泮单光子发射计算机断层扫描(SPECT)对大脑中中枢苯二氮䓬受体结合潜能进行定量分析:技术与观察者间的变异性

Quantification of central benzodiazepine receptor binding potential in the brain with 123I-iomazenil SPECT: technical and interobserver variability.

作者信息

Verhoeff N P, Erbas B, Kapucu O, Busemann Sokole E, Blok H, Van Royen E A

机构信息

Department of Nuclear Medicine, Academic Medical Center, Amsterdam, The Netherlands.

出版信息

Nucl Med Commun. 1993 Aug;14(8):634-43. doi: 10.1097/00006231-199308000-00003.

Abstract

Factors contributing to the quantification of the central benzodiazepine receptor binding potential in vivo using 123I-iomazenil and single photon emission computed tomography (SPECT) were analysed in phantom studies and in volunteers. SPECT was performed with the SME810 multidetector system. The Hoffman three-dimensional brain phantom was used to investigate linearity and reproducibility of SPECT results using different 123I activity concentrations. Dynamic and multislice SPECT scans were performed in nine healthy volunteers between 0 and 270 min after injection of 110 MBq 123I-iomazenil. Displacement studies were performed in three volunteers using 1 mg flumazenil administered intravenously (i.v.) at 250 min postinjection (p.i.). Regions of interest (ROIs) over the cerebral cortex and various subcortical nuclei were either fixed areas based on a stereotaxic brain atlas, or manually drawn, based on 60% isocontour lines. Interobserver variability of the ROI data and ratios derived from those data were estimated for two independent observers. The brain phantom measurements showed linearity with respect to 123I concentration and good reproducibility. The interobserver study showed a reasonable interobserver reliability for the large fixed ROIs. The displacement study showed about 50% displacement in all ROIs. Concentrations of 123I-iomazenil in the brain can be measured reliably with the SME810. Large fixed ROIs based on a stereotaxic atlas may be used reliably for quantification. The level of 4 cm above the cantomeatal line does not provide a useful reference region with a very low central benzodiazepine receptor binding potential.

摘要

在体模研究和志愿者中分析了使用123I-碘美西尼和单光子发射计算机断层扫描(SPECT)在体内定量中枢苯二氮䓬受体结合潜能的相关因素。使用SME810多探测器系统进行SPECT检查。利用霍夫曼三维脑体模,使用不同的123I活度浓度研究SPECT结果的线性和可重复性。在9名健康志愿者注射110 MBq 123I-碘美西尼后的0至270分钟内进行动态和多层SPECT扫描。在3名志愿者中进行了置换研究,在注射后250分钟静脉注射(i.v.)1 mg氟马西尼。大脑皮质和各个皮质下核的感兴趣区(ROI)要么是基于立体定向脑图谱的固定区域,要么是基于60%等轮廓线手动绘制的。对两名独立观察者估计了ROI数据和从这些数据得出的比率的观察者间变异性。脑体模测量显示出与123I浓度相关的线性和良好的可重复性。观察者间研究表明,对于大的固定ROI,观察者间具有合理的可靠性。置换研究显示所有ROI中的置换率约为50%。使用SME810可以可靠地测量脑中123I-碘美西尼的浓度。基于立体定向图谱的大的固定ROI可可靠地用于定量。外耳道中心线上方4 cm的水平不能提供具有非常低的中枢苯二氮䓬受体结合潜能的有用参考区域。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验