Mackey N A, Ilsley A H, Owen H, Plummer J L
Department of Anaesthesia and Intensive Care, Flinders Medical Centre, Bedford Park, South Australia.
Anaesth Intensive Care. 1993 Apr;21(2):204-7. doi: 10.1177/0310057X9302100213.
A low-cost, disposable patient-controlled analgesia device was tested using a computer-controlled, automated testing station designed to simulate clinical use. Five devices were tested and delivered a mean bolus dose size of 95% of nominal, although one device achieved only 85% of nominal. Delivery was lowest immediately following an eight-hour interval during which no demands were made. Further development and evaluation of this interesting device is warranted.