Three hearing aid selection procedures were compared to determine if any one was superior in producing prescribed real-ear insertion gain. For each of three subject groups, 12 in-the-ear style hearing aids with Class D circuitry and similar dispenser controls were ordered from one of three manufacturers. Subject groups were classified based on the type of information included on the hearing aid order form: (1) the subject's audiogram, (2) a three-part matrix specifying the desired maximum output, full-on gain, and frequency response slope of the hearing aid, or (3) the desired 2-cc coupler full-in grain of the hearing aid, based on real-ear coupler difference (RECD) measurements. Following electroacoustic adjustments aimed at approximating a commonly used target insertion gain formula, results revealed no significant differences among any of the three selection procedures with respect to obtaining acceptable insertion gain values.