Hellbusch L C
Department of Surgery (Neurosurgery), University of Nebraska Medical Center, Omaha 68114, USA.
Childs Nerv Syst. 1996 Mar;12(3):142-4. doi: 10.1007/BF00266816.
Three cases of shunt valve fracture with separation are reported. Fracture with separation of the shunt valve of a cerebrospinal fluid shunt is an infrequent occurrence. Because a portion of the shunt valve is radiolucent on a shunt survey radiograph, this problem is easily overlooked if it occurs. The diagnosis can be made by comparing old and new shunt survey radiographs and noting the increased distance between radiopaque portions of the shunt on the new x-ray films.
报告了三例分流阀骨折伴分离的病例。脑脊液分流器的分流阀骨折伴分离是一种罕见的情况。由于分流阀的一部分在分流造影X光片上是透光的,如果出现这种问题很容易被忽视。通过比较新旧分流造影X光片,并注意新X光片上分流器不透光部分之间距离的增加,可做出诊断。