Katmeh M F, Aherne G W, Stevenson D
Robens Institute, University of Surrey, Guildford, UK.
Analyst. 1996 Mar;121(3):329-32. doi: 10.1039/an9962100329.
The feasibility of using an enhanced chemiluminescent immunoassay for screening water samples was adapted to a Dynatech Microlite camera luminometer and assessed to provide a semiquantitative assay based on a photographic record of the luminescent end-point. An enhanced chemiluminescent immunoassay of the herbicide chlortoluron was chosen in this instance as an example, although other pesticide compounds could equally well have been used provided that suitable antisera were available. This luminescent assay has shown considerable potential by providing a rapid, simple and portable means of monitoring multiple water samples for the presence of chlortoluron. The assay was able to identify samples containing the herbicide at or above the European limit for individual pesticides in drinking water [the EU Maximum Admissible Concentration (MAC) = 0.1 microgram l-1]. A 100% accuracy was obtained in the analysis of samples containing chlortoluron at concentrations above the range of 0.07-0.12 microgram l-1; however, for concentrations within this range the results were 65% accurate. Consequently, the method could greatly facilitate increased monitoring for the presence of chlortoluron in water supplies and in other areas of environmental analysis.
将用于水样筛查的增强化学发光免疫分析法适配到Dynatech Microlite相机发光计上,并基于发光终点的照片记录进行评估,以提供一种半定量分析方法。在本例中,选择了除草剂绿麦隆的增强化学发光免疫分析法作为示例,不过只要有合适的抗血清,其他农药化合物同样可以使用。这种发光分析法通过提供一种快速、简单且便携的手段来监测多个水样中绿麦隆的存在,已显示出相当大的潜力。该分析法能够识别出含有除草剂且达到或超过饮用水中单个农药的欧洲限量标准(欧盟最大允许浓度(MAC)= 0.1微克/升)的水样。在分析浓度高于0.07 - 0.12微克/升范围的含有绿麦隆的水样时,准确率达到100%;然而,对于该浓度范围内的水样,结果的准确率为65%。因此,该方法可极大地促进对供水及其他环境分析领域中绿麦隆存在情况的监测。