测量误差对M型超声心动图测定左心室质量的影响。

Influence of measurement inaccuracies on determination of left ventricular mass by M mode echocardiography.

作者信息

Stöllberger C, Holländer I, Dimitrov L, Slany J

机构信息

Zweite Medizinische Abteilung der Krankenanstalt Rudolfstiftung, Vienna, Austria.

出版信息

Heart. 1996 Mar;75(3):312-3. doi: 10.1136/hrt.75.3.312.

Abstract

OBJECTIVE

To determine to what extent inaccuracies in measuring the end diastolic diameter of the left ventricle, the interventricular septum, and the posterior wall, by M mode echocardiography influence the left ventricular mass calculated by the Devereux's formula.

DESIGN

Mathematical model.

RESULTS

Relatively small measurement inaccuracies such as 5%, which are known to be inherent in the echocardiographic method, will result in "changes" of left ventricular mass in the range of 8% to 15%. This is equivalent to expected changes in left ventricular mass seen during treatment.

CONCLUSIONS

The use of Devereux's formula to calculate left ventricular mass is limited by measurement inaccuracies in individual patients.

摘要

目的

确定M型超声心动图测量左心室舒张末期内径、室间隔和后壁时的不准确性在多大程度上影响通过德弗罗公式计算的左心室质量。

设计

数学模型。

结果

相对较小的测量不准确性,如已知超声心动图方法中固有的5%,将导致左心室质量在8%至15%范围内“变化”。这相当于治疗期间左心室质量的预期变化。

结论

使用德弗罗公式计算左心室质量受到个体患者测量不准确性的限制。

相似文献

5
A comparison of MRI and echocardiography in hypertrophic cardiomyopathy.
Br J Radiol. 1999 Mar;72(855):258-64. doi: 10.1259/bjr.72.855.10396215.
7
Maternal left ventricular mass and diastolic function during pregnancy.
Ultrasound Obstet Gynecol. 2001 Nov;18(5):460-6. doi: 10.1046/j.0960-7692.2001.00573.x.

本文引用的文献

3
Prognostic implications of echocardiographically determined left ventricular mass in the Framingham Heart Study.
N Engl J Med. 1990 May 31;322(22):1561-6. doi: 10.1056/NEJM199005313222203.
6
Inter-study variability in left ventricular mass measurement. Comparison between M-mode echography and MRI.
Eur Heart J. 1992 Aug;13(8):1011-9. doi: 10.1093/oxfordjournals.eurheartj.a060307.
7
Reproducibility of echocardiographic estimates of left ventricular dimensions.
Br Heart J. 1977 Apr;39(4):390-8. doi: 10.1136/hrt.39.4.390.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验