Chen S H, Yang Z Y, Wu H L, Kou H S, Lin S J
Graduate Institute of Pharmaceutical Sciences, Kaohsiung Medical College, Taiwan, Republic of China.
J Anal Toxicol. 1996 Jan-Feb;20(1):38-42. doi: 10.1093/jat/20.1.38.
A simple and sensitive high-performance liquid chromatographic (HPLC) method was established for the trace determination of thiocyanate anion as a fluorogenic derivative. The method is based on the chemical derivatization of thiocyanate anion with 3-bromomethyl-7-methoxy-1,4-benzoxazin-2-one. The resulting derivative was separated by a Nova-Pak C18 reversed-phase column. Optimization conditions for the derivatization of thiocyanate anion were investigated by HPLC with fluorimetric detection. The linear range for the quantitation of thiocyanate anion was 1-0.05 nmol in 0.1 mL of sample; the detection limit (with a signal-to-noise ratio of 5) of a 20-microL injected aliquot was approximately 3.3 +/- 1.2 fmol. Application of the method to the analysis of thiocyanate anion in saliva and plasma proved to be feasible.